Re: [AMMRL] Unexpected peaks of 19F NMR from CF3CH2OH

From: GOBET Mallory <mallory.gobet_at_arkema.com>
Date: Wed, 12 Jul 2023 18:23:20 +0000

Hi Weidong. There is a 12C/13C isotope effect in 19F NMR. also observed
when 19F is coupling with 35/37Cl and 10/11B

http://u-of-o-nmr-facility.blogspot.com/2012/10/isotope-effects-and-19-f-13-c-hmqc.html


Mallory Gobet
Analytical and Systems Research Group
Arkema Inc.

From: main_at_ammrl.groups.io <main_at_ammrl.groups.io> On Behalf Of Weidong Hu via groups.io
Sent: Wednesday, July 12, 2023 1:33 PM
To: main_at_ammrl.groups.io
Subject: [AMMRL] Unexpected peaks of 19F NMR from CF3CH2OH

Hi All,

I acquired 1D-19F NMR of CF3CH2OH dissolved in CDCl3/CD3OD(2:1 v/v) on CP TCI
700 H&F-C/N-D probe (without 1H decoupling), and the spectrum is very confusing.

In addition to main triplet peaks, there are three minor triplet peaks (about
0.5% of main triplet). The positions of minor peaks are not symmetric to the
main triplet, and thus should not come from natural abundance of 13C-19F J
coupling. It is reasonable to see four minor peaks generated from one and
two-bond J couplings between 13C-19F, but they should be symmetric to the
main peak. I don't know how to explain the observed spectrum. Any suggestions
as what is going on? The spectrum is attached below.

Thank you,

Weidong

Dr. Weidong Hu
Department of Immunology & Theranostics
Beckman Research Institute
1450 E. Duarte Rd., Duarte, CA 91010
Tel: 626 256 4673, Ext: 63416
Fax: 626 301 8186




________________________________

-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#306): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/306__;!!PvDODwlR4mBZyAb0!TrnC-XAZYblA5sVFxLVuLTimvz-bNfY7=
pdD_MciuCyGqJ2sNzXujXnf4eLd0xlka5QakaEEDipseFS7w-WPGqeo$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/100104757=
/7559972__;!!PvDODwlR4mBZyAb0!TrnC-XAZYblA5sVFxLVuLTimvz-bNfY7pdD_MciuCyGqJ=
2sNzXujXnf4eLd0xlka5QakaEEDipseFS7wfw_1g_M$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Jul 12 2023 - 12:01:46 MST

This archive was generated by hypermail 2.4.0 : Wed Oct 25 2023 - 14:43:02 MST