Hello Again All:
Here is what I discovered regarding the extra pulse lines labeled
"AUX1" and "AUX2" on the Mercury consoles. Don Eldred suggested
trying "aux1on();" and "aux12off();" - sure enough, these do work,
however "dps" gives a segmentation violation. On a hunch, I tried
putting in "#ifdef DPS" and "#endif" statements on either side of
these statements in the pulse program, and that solved the dps
problem. I also note that "aux2on():" can be used to turn on the
"AUX2" port, with aux12off(); also turning that one off. It appears
no command exists to turn off just one of the AUX ports as "aux1off
()" and "aux2off()" won't compile (symbols don't exist). This is no
problem for me at this point as I just need one of the ports to gate
an amplifier.
Thanks again to all who replied.
Tom Pratum