Summary: Apodization of NOESY

From: Edward T. Chainani <etch_at_ginto.chem.admu.edu.ph>
Date: Mon, 10 Jul 2000 10:16:14 +0800 (PHT)

Thanks to all who responded to my question below: Deane MacIntyre, Mo
Latif, Hermann Kalchhauser.

To sum up: sine squared shifted 90 degrees for magnitude or phase
sensitive NOESY. Another is qsine shifted 90 for phase sensitive, and
sine unshifted for magnitude. Will try both!

> In light of the recent post regarding apodization functions for HMQC,
> etc., I'd like to know the proper functions for a NOESY experiment. We
> have JEOL Lambda 400, which maybe similar to the JEOL Eclipse+.

------------------------------------------------------------------------------
Edward T. Chainani Nuclear Magnetic Resonance Laboratory Manager

             National Chemistry Instrumentation Center (NCIC)

Ateneo de Manila University phone: (632) 426-6001 loc. 5619
Loyola Heights, Quezon City e-mail: etch_at_ginto.chem.admu.edu.ph
Philippines FAX: (632) 426-5985
------------------------------------------------------------------------------
Received on Mon Jul 10 2000 - 10:18:12 MST