Re: [AMMRL] helium plumbing on vintage Bruker magnets

From: Michael Groves via groups.io <mgrovesnmr=yahoo.com_at_groups.io>
Date: Tue, 27 Jan 2026 22:09:53 +0000 (UTC)

Hi David,

I don't know what the D number of your magnet is--you can check that on the
base plate.  Here's what I have regarding replacing the existing manifold
depending on the D number of your magnet.

Cheers,Mike


On Tuesday, January 27, 2026 at 01:30:11 PM MST, Vander Velde, David via groups.io wrote:

 Hi,
I am now involved in managing a Bruker Ultrashield magnet (which I had
previously done decades ago) after a period of managing Ascend magnets.
Prior to the Ultrashield series, every magnet I worked with (Oxford or
Spectrospin) had a cross piece connecting however many helium stacks there
were, with a small one way valve on the single vent.. The Ultrashields came
without this piece, with just Tygon tubing connecting the stacks to a similar
one way valve. With the Ascend series, the cross piece came back, with a
much more robust one way valve which can stay in place during helium fills.
(pictures attached)

I took care of Ultrashields for quite a while without understanding why the
cross piece went away, or what the really optimal way was to vent the cold
gas released during a helium fill. The two stacks on the mid field magnets
behave very differently--the one with the fill port and the helium sensor
appeared to be wide open to exiting helium, whereas the other stack was
extensively set up with baffles or other internal structures which impeded
the helium. 

I have a very strong preference for filling magnets that feature the cross
piece, where how much helium is exiting through which stack is transparent
to me.

My questions are, has anyone modified their UltraShield with the Ascend style
cross piece? Alternatively, does anyone know of reasons why that would be
a bad idea?

--
Dave VanderVelde


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3163): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3163__;!!PvDODwlR4mBZyAb0!UrztPMSHe5RQic5OP6XlarW8CgWyiv=
inYNzhafrGX5VqkwzImh-qb3fGLF46cxdc_8CpSHtMTbbxeYlXoff0tn4Kv_Ag$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/117496151=
/7559972__;!!PvDODwlR4mBZyAb0!UrztPMSHe5RQic5OP6XlarW8CgWyivinYNzhafrGX5Vqk=
wzImh-qb3fGLF46cxdc_8CpSHtMTbbxeYlXoff0tspDjKkQ$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-






1769551753876blob.jpg
(image/png attachment: 1769551753876blob.jpg)

Received on Tue Jan 27 2026 - 14:25:21 MST

This archive was generated by hypermail 2.4.0 : Mon Feb 09 2026 - 16:54:28 MST