Re: [AMMRL] Bruker Avance III : No communication between the SGUs and Amplifiers

From: Igor Moudrakovski <i.moudrakovski_at_fkf.mpg.de>
Date: Tue, 27 Jan 2026 17:31:46 +0100

Hi Mike,

Thank you for your comments.

The service webpage for each amp in the the Channel Information has a a
column Input Num that says Via router, but each of 3 SGUs is wired
directly to one of the three amps (SGU(1) RF OUT J2 ->> BLAX1000 IN
1, SGU(2) RF OUT J2 ->> BLAH300 IN 1, SGU(3) RF OUT J2 ->> BLAX300 IN
1). I am a bit lost on this routing.

We did already try a different Ethernet switch without success. We also
checked one by one all the RF cables  between the SGUs and the amps, as
well as the LAN cables between the Ethernet switch and the amps.

Regards,

Igor

  On 26/01/2026 18:06, Michael Groves via groups.io wrote:
> Hi Igor,
>
> Most frequently when we see this error it's just because something
> didn't get powered on in the right order so the amplifier just needs
> to get a new IP address assigned so that they can communicate again. 
> Normally you can resolve this by simply power cycling the amplifiers,
> waiting for them to boot up, and running cf again. This might not be
> the cause of your errors.  There are a couple of other things that you
> can check.
>
> When you open up ha and see the amplifiers listed, can you open the
> service webpage?  And do these amplifiers have input routers or do
> they just have one input?  You might check the ethernet switch in the
> console just to see if you've got a couple of bad ports or some other
> communication failure.  You might try a different input to the
> amplifier--we have seen cases where one or more of the router inputs
> fails so when the SGU pulses during that part of cf, no pulse gets to
> the amplifier and you get the "no connection detected" message.  I
> assume there's not a router in the AQS rack that routes the SGU
> signals to different amplifiers.  If there was one of these I might
> try bypassing that as well. How do the cables look between the SGUs
> and amplifiers? Normally those don't fail but it might be worth
> disconnecting them at both ends and seeing if they're in good shape,
> and installed correctly. (note--I would expect that to be the case but
> it doesn't hurt to be thorough.)  Since you don't have a valid
> configuration you can't pulse on the amplifiers but if you were to use
> nmr_restore you might be able to get back a valid configuration and
> just try pulsing on the amps and see what inputs and outputs are
> present.  If you've got an oscilloscope you can have a look at the
> input and output pulses as well as the blanking signal.
>
> Good luck,
> Mike
>
>
>
> On Monday, January 26, 2026 at 09:08:52 AM MST, Igor Moudrakovski wrote:
>
>
> Hello Everyone,
>
> I have a Bruker Avance III 400 MHz solid-state system circa 2010
> (Cento5.11, Topspin 3.6.2), which developed over time a communication
> problem between the SGUs and the amplifiers. The CF proceeds seemingly
> without any problem, only to end up with an error message
> ‘No connection between SGUs and amplifiers found! Acquisition will not
> be possible. The wiring may be incorrect, or the emergency stop
> signal may be active. Please check the hardware and do ‘cf’ again.’
>
> No visible problems with the hardware; all LEDs in front and back are
> green, no warning/flashing lights, and no connections have changed since
> the instrument was functioning. The ‘ha’ reveals no obvious problems
> either.
>
> Last year, Bruker responded to the problem by replacing all the fans and
> filters in AQS/3, as well as the amplifiers. That resulted in about two
> months of stable work, but apparently didn’t solve the root problem. Now
> the instrument is back to the state of ‘No connection between SGUs and
> amplifiers…’ Numerous reboots didn’t move the problem.
>
> I would be very grateful to hear all your suggestions for resolving the
> issue.
> Thank you and best regards,
>
> Igor
> --
> Igor Moudrakovski, PhD
> Max-Planck Institute for Solid State Research
> Heisenbergstr. 1, 70569 Stuttgart, Germany
> Office 711-689-1963
> 400 NMR 711-689-1410
> 600 NMR 711-689-1054
>
>


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3160): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3160__;!!PvDODwlR4mBZyAb0!WAXiJvB0x9ySIXQIJKCEh5qSvaDwk8=
K_bfCQ86fxwX5hZV6imp2h8GtV3w1_GSWz_u-qw1kdVVInVW17-Zpo1eZ9Uxp5$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/117471018=
/7559972__;!!PvDODwlR4mBZyAb0!WAXiJvB0x9ySIXQIJKCEh5qSvaDwk8K_bfCQ86fxwX5hZ=
V6imp2h8GtV3w1_GSWz_u-qw1kdVVInVW17-Zpo1ScjZVuP$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Jan 27 2026 - 08:34:12 MST

This archive was generated by hypermail 2.4.0 : Mon Feb 09 2026 - 16:54:28 MST