Re: [AMMRL] Topshim with parameters and Autoshim turned on

From: Anklin, Clemens via groups.io <clemens.anklin=bruker.com_at_groups.io>
Date: Mon, 19 Jan 2026 20:22:12 +0000

Hi Andre,

set the shim program in the ICONNMR configurator to
"TOPSHIM_WITH_AUTOSHIM convomp tuneay tuneax"

this should do the job

the shims to use during autoshim are as usual activated from the BSMS panel.

Best regards
Clemens


On 1/19/2026 12:56 PM, Andre Dallmann via groups.io wrote:
>
> Dear all,
>
> has anyone a quick way to get ICON to make "topshim convcomp tunay
> tuneax" and afterwafrds turning on AUTOSHIM (ideally with z1z2xy)?
>
> Sure this must be easy, but somehow cannot find the solution...
>
> Thanks and best, André
>
> --
> Dr. André Dallmann
> Leiter der NMR-Abteilung
>
> Email: andre.dallmann_at_chemie.hu-berlin.de
> Raum: 0'103
> Telefon: 82952
> Handy: 01797657855
> ORCID-ID: https://orcid.org/0000-0002-0141-2635
> Webseite:
> https://www.chemie.hu-berlin.de/en/forschung-en/nmr-spectroscopy/nmr-spectroscopy?set_language=en
>
> Adresse:
> Humboldt Universität zu Berlin
> Institut für Chemie
> AG NMR
> Brook-Taylor-Str. 2
> 12489 Berlin
> Deutschland
>
>
>
--
==========================
==========================
====================
Clemens Anklin Ph.D. Vice President
Bruker BioSpin NMR Applications & Training
15 Fortune Drive NEW Phone: 978-313-5NMR(5667)
Billerica MA 01821 bluesky: _at_canklin.bsky.social
web: http://www.bruker.com
e-mail: clemens.anklin_at_bruker.com
==========================
==========================
====================




-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3134): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3134__;!!PvDODwlR4mBZyAb0!X5rCDGcZ2WeKd5u5H-Cgetipyh56jh=
MvHEcx-XpHp0Kc_ys62bGos77UArEKT8HTD-Wq3g3L1ArJQW8Fz9B5yMxGHFPU$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/117349140=
/7559972__;!!PvDODwlR4mBZyAb0!X5rCDGcZ2WeKd5u5H-Cgetipyh56jhMvHEcx-XpHp0Kc_=
ys62bGos77UArEKT8HTD-Wq3g3L1ArJQW8Fz9B5yN3ObW5t$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-


Received on Mon Jan 19 2026 - 18:22:02 MST

This archive was generated by hypermail 2.4.0 : Mon Feb 09 2026 - 16:54:28 MST