Re: [AMMRL] Real Natural Rubber Tubing

From: Joseph Dumais via groups.io <josephdumais=boisestate.edu_at_groups.io>
Date: Thu, 8 Jan 2026 13:31:28 -0700

 have been using copper tubing (the flexible refrigerator type) with the
black foam insulation over it and then a short piece of rubber tubing,
about 18 inches, at the magnet end. The copper portion mounts to the LN2
tank with a flare nut. I have done this for many years. If the tubing
shatters you are at the least 5 feet away from the break. Of course I do
wear safety glasses and gloves too.

I have not worked it out but I suspect that the copper tubing has a lower
heat capacity than the rubber tubing since the transfers seem to be a bit
quicker. The only issue is that you need to place the tank fairly well and
I work to insure that as the rubber tubing starts to freeze that the bend
is a smooth radius.

I have had a few breaks in the rubber tubing but I have always been a safe
distance away from the breaks.


*******************************
Joe Dumais, Ph.D. (retired)
Associate Research Professor-Emeritus
Nuclear Magnetic Resonance Spectroscopist
Boise State University
College of Arts and Sciences
Department of Chemistry and Biochemistry
1910 University Drive, Boise, Idaho 83725-1520
(208) 426 4913
https://www.boisestate.edu/chemistry/
*******************************

“The best that most of us can hope to achieve in physics is simply to misunderstand at a deeper level.” -*Wolfgang Pauli*

Sent from my MS Windows PC



On Wed, Jan 7, 2026 at 9:04 AM Raphael Stoll via groups.io wrote:

> Dear fellow colleagues,
>
> developed by one of our former co-workers, we have been using the
> following insulated hose for several years now:
>
> https://nmrtools.com/product/nmr-ln2-hose/
>
> It combines safety with efficient nitrogen transfer and works a treat.
>
> Cheers,
> Raphael
>
>


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3114): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3114__;!!PvDODwlR4mBZyAb0!VjwyEnnJEgFpvcpn5ubs8WAwmzYNkW=
dj_hyjXHRIbRfLV2vquWwefl0oeYgdg1DWJNIloAwjgWEBMm2dfkhKL7OKsOm6$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/116902195=
/7559972__;!!PvDODwlR4mBZyAb0!VjwyEnnJEgFpvcpn5ubs8WAwmzYNkWdj_hyjXHRIbRfLV=
2vquWwefl0oeYgdg1DWJNIloAwjgWEBMm2dfkhKLw8Qa5OP$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Jan 08 2026 - 18:25:53 MST

This archive was generated by hypermail 2.4.0 : Mon Feb 09 2026 - 16:54:28 MST