[AMMRL] Agilent AS7600 error

From: vbegue via groups.io <vbegue=eastman.com_at_groups.io>
Date: Mon, 06 Jan 2025 07:53:51 -0800

We've been getting insertion errors on our Agilent DD2 system lately. 
Errors have been intermittent and broad in where they occurred.  We've
finally been able to trace it back to wear on the gripper fingers of our
AS7600 sample changer.  The aluminum on the fingers has worn down to the
point where samples are not gripped reliably and are intermittently not
picked up or dropped mid transport.  Our system has been running 24/7
for roughly 14 years.  We applied adhesive foam craft stickers cut to
size at the 4 contact points of the gripper and the robot is transferring
reliably again.


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1949): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1949__;!!PvDODwlR4mBZyAb0!Xo5TZGGgoUKc1KIJ-0emhiu2MOa8bW=
AXM3IlER2-uI85jqDixYojsg3f0x0a606sVEpS8rSdO-2lAxnSpXK0xKEohzOO$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/110459841=
/7559972__;!!PvDODwlR4mBZyAb0!Xo5TZGGgoUKc1KIJ-0emhiu2MOa8bWAXM3IlER2-uI85j=
qDixYojsg3f0x0a606sVEpS8rSdO-2lAxnSpXK0xOwlzNl6$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon Jan 06 2025 - 11:43:12 MST

This archive was generated by hypermail 2.4.0 : Sat Feb 01 2025 - 17:12:42 MST