[AMMRL] Shimming issues related to oscillating lock level

From: Kenneth Sharp-Knott <kknott_at_vt.edu>
Date: Thu, 11 Jan 2024 15:13:34 -0500

*Topline: Issue still not resolved - D20 samples unusable - CDCl3 looks normal*

I'm now going to spam you with the information I've collected over the week
based on all of the excellent suggestions from the AMMRL users!

   1. This doesn't seem to be an issue with shimming, but rather something
   impacting both the linewidth and chemical shifts of D2O samples in
   particular.
   2. I cannot correct the lineshape manually. It is not correctable with
   tcorr.
   3. Spinning vs Non-Spinning - Improved lineshape standard. Strange
   change to D2O sucrose sample. PDF attached
   4. Menthol in CDCl3 lineshape is excellent
   5. Possibly very very small chemical shift changes
   6. Menthyl Anthranilate is DMSO is good
   7. The probe coil temp is stable at 27
   8. The VT gas flow is stable
   9. The VT heater power is stable (no fluctuation beyond 0.1% heater
   power)
   10. The VT sample temp is stable (no fluctuation beyond 0.1 Celsius)
   11. Shim Coil Temp is stable at 300K
   12. In order to eliminate the possibility of issue related to our house
   air, I've tried using both our usual house air and boil off liquid high
   pressure N2 - no difference
   13. Co59 samples chemical shift is stable for 10 minutes (every 10
   seconds)
   14. Co59 sample - there was some change in phase.
   15. Room Temperature BBO probe does not exhibit any issues -
   shimming/lineshape is perfect across all solvents
   16. Attached borescope of the probe shows there was some gunk on the
   sample cavity walls. I cleaned it - no change
   17. After returning the Prodigy to the magnet (24 hour pump down) - no
   change
   18. The lock level oscillation is observed with acetone - not D2O
   19. The Prodigy platform is overdue for service (scheduled this month),
   but there are no indications it is having trouble maintaining vacuum or
   dewar pressure.

I was pretty certain there was some kind of temperature gradient through
the sample, but the Co59 shift being rock solid has me doubting that. I'm
pretty much at a total loss.

Sending it to the AMMRL team for more suggestions

Considering opening it up for organic solvent use until the engineer
services the Prodigy platform.

--
Ken Sharp-Knott
Manager of Analytical Services and the NMR Facility
Department of Chemistry
Virginia Tech

(540)267-6502 (Cell)
(540)231-0885 (Office)


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#922): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/922__;!!PvDODwlR4mBZyAb0!WzJqH_e8O6y02srcJgZxWNS8nRFqBEkz=
AnMKyM5YXAmBHdga_Pxa4B5DRiJMuGT-iJ2C2Ea9K82Gmg$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/103602621=
/7559972__;!!PvDODwlR4mBZyAb0!WzJqH_e8O6y02srcJgZxWNS8nRFqBEkzAnMKyM5YXAmBH=
dga_Pxa4B5DRiJMuGT-iJ2C2EbTMWoDEA$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Jan 11 2024 - 12:16:37 MST

This archive was generated by hypermail 2.4.0 : Thu Feb 01 2024 - 14:56:24 MST