RE: solid to solution phase change time

From: David Kiemle <dkiemle_at_syr.edu>
Date: Tue, 7 Jan 2003 12:22:31 -0500

We have a Bruker AVANCE 600 MHz system with a standard bore magnet. I
routinely switch the system from CPMAS, HRMAS and Standard Liquid
probes. It is as simple as a standard probe change over with the
exception that you must run KBr, Adamantine, and glycine. This typically
takes about 1/2 hour. Shims never change, and with Linear amplifiers
pulse widths do not change much either. To switch back takes minutes.

Dave Kiemle
SUNY ESF