Dear Friends:
I posted a message about our Bruker ARX300 last October. We had a lot of
problems with it. One of the problems was the S/N decreased when the NS was
increased.
I would like to thank the following friends for their kind comments and
suggestions:
E. Brooks, R. D. Scott, M. Stchedroff, S. Macura, C. Rithner, M. Leister Jr.,
Redwine David, K. A. Smith, D. Hasha, W. Kurt, P. Xu, L. Gregory, R. Tschudin,
J. Breeyear and D. V. Velde.
Bruker engineers in Houston worked very hard on our ARX300. It turned out that
the MCI board might cause the S/N problem. We changed 3-4 MCI boards. The final
one solved our S/N problem.
Bruker engineers also solved the other problems on the ARX300 for us. Our Bruker
ARX300 is up and running.
Thanks again for your and Bruker engineers' help.
See you in Orlando.