Re: [AMMRL] BSMS Error: L-TRX 2H Amplifier when shimming

From: Anklin, Clemens via groups.io <clemens.anklin=bruker.com_at_groups.io>
Date: Sat, 21 Feb 2026 14:28:42 +0000

Hello Vernon

DMF is a tricky solvent to run gradient shimming on. The signal picked for
shimming is one of the methyl groups which are less than 0.2 ppm apart. In
order to only excite this one signal a selective pulse is applied to it. The
duration of this pulse is given by the parameter SELWID in the solvent specific
parameters. For DMF this is set to 0.15 ppm

It looks like under these conditions the pulse length gets to be over 1 second
and the deuterium amplifier does not like this.

You can specify other parameters to be used with a solvent in automation.
You can try the formyl signal at 8.61 ppm. It will be less intensive but
should still work. The command would then look somethig like "topshim 2h
o1p=8.61 selwid=1.0"

Or you can just try "topshim selwid=0.2" but there is a risk that it catches
too much of the other methyl group and the results are going to be bad.

With best regards
Clemens

On 2/20/2026 3:55 PM, Vernon M Russell via groups.io wrote:

I don't run DMF experiments very often but today I received the following
error when topshim was running in automation mode on a sample dissolved in
DMF.

BSMS Error:
L-TRX 2H Amplifier: RF Pulse Length or Duty Cycle violation! Please check
your Pulse Program. Max Duty Cycle: 10%, Max Pulse Length: 1000ms

I have not made any changes to the standard ICON-NMR settings for Topshim
nor the DMF listing in edlock. The sample eventually shims okay and data
collected for the sample. Any help would be appreciated.

Thanks,


Vernon Russell

vrussell_at_wlgore.com
T +1-410-506-3852
W. L. Gore & Associates, Inc.
Together, improving life
501 Vieve's Way, P.O. Box 1320
Elkton, MD 21922-1320
USA
gore.com



For information about our privacy practices, see our Privacy Notice
https://www.gore.com/privacy

This email may contain trade secrets or privileged, undisclosed or otherwise
confidential information. If you have received this email in error, you are
hereby notified that any review, copying or distribution of it is strictly
prohibited. Please inform us immediately and destroy the original transmittal.
Thank you for your cooperation.



--
==========================
==========================
====================
Clemens Anklin Ph.D. Vice President
Bruker BioSpin NMR Applications & Training
15 Fortune Drive NEW Phone: 978-313-5NMR(5667)
Billerica MA 01821 bluesky: _at_canklin.bsky.social
web: http://www.bruker.com
e-mail: clemens.anklin_at_bruker.com<
==========================
==========================
====================



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3252): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3252__;!!PvDODwlR4mBZyAb0!VV9kZay872TFSAbt2ongUEIvSSD727=
pAnL9Lq_IOBxa7pDAYoPGaHfTUtIYUF9nXwR65Jf9n97WO1Oj1Ypx-mHOfm_0p$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/117917230=
/7559972__;!!PvDODwlR4mBZyAb0!VV9kZay872TFSAbt2ongUEIvSSD727pAnL9Lq_IOBxa7p=
DAYoPGaHfTUtIYUF9nXwR65Jf9n97WO1Oj1Ypx-mEgZP8fw$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Feb 25 2026 - 10:54:54 MST

This archive was generated by hypermail 2.4.0 : Sun Mar 01 2026 - 13:47:31 MST