Re: [AMMRL] Any brave intrepid explorers installed Topspin 5.0 yet?

From: Matthias Brandl via groups.io <matthias.brandl=tum.de_at_groups.io>
Date: Tue, 17 Feb 2026 11:34:04 +0100

Hi all,

Just a small note: I just tested apbk in the Linux processing-only
version, and it behaves as advertised (no integration without options,
apbk -w does integrate). So Geoffrey's issue seems to likely be Windows
related.

Cheers,
Matthias


On 12/02/2026 18:16, Akien, Geoffrey via groups.io wrote:
> Hi folks,
>
> Got 5.0.0 working on our Neo on Windows 11 eventually. Predictably I did
> something similar to Ken during installation and got caught out by the
> passwords for nmradministrator – initial passwords are generated and
> stored in plaintext in (Windows 11 at least) C:
> \Bruker\Credentials_YYYY_MM_DD.txt. For those on the fence, its not that
> different to 4.5.0 really.
>
> I’m seeing some weird text rendering issues on Windows 11 to the point
> that the standard text editor (edau etc.) is unusable. We’re on standard
> font sizes and 100% scaling in Windows, so really not sure what’s going
> on there. I have a vague memory about there being a text file to change
> to fix this, but I had a look through my notes and previous release
> letters and couldn’t find anything. Maybe someone can remember better
> than me?
>
> “New” apbk was supposed to stop integrating spectra by default, but
> unfortunately still is, and also now ignores -n so you can’t even turn
> it off anymore. Which is shame, because of the lot of the new options
> look really useful.
>
> For figuring out the new components in TS 5, the manuals for IAM and
> SpinPilot are pretty reasonable starting points.
>
> Whether to switch SpinPilot is another matter entirely though. There are
> currently no tools for migrating IconNMR configurations into SpinPilot,
> so if you have lots of users, solvents and parameter sets, this is not
> going to be a fun experience at all. At least to my eyes it isn’t
> feature complete with IconNMR - from the manual: “The initial version of
> SpinPilot does not yet include the full range of NMR automation functions.”.
>
> Given IconNMR will continue to exist, and many of us will be supporting
> instruments stuck on IconNMR for perpetuity, I’m not sure its worth the
> pain of migrating just yet, not to mention the additional training to
> keep users happy on both systems. If we only had Neo’s the calculus
> would be a little different, but not sure how many labs are that fortunate.
>
> Is anyone aware of timelines for IconNMR deprecation and/or SpinPilot
> feature addition?
>
> Thanks
>
> Geoff
>
> *From:*main_at_ammrl.groups.io <main_at_ammrl.groups.io> *On Behalf Of
> *Kenneth Sharp-Knott via groups.io
> *Sent:* 11 February 2026 20:20
> *To:* main_at_ammrl.groups.io
> *Subject:* [External] Re: [AMMRL] Any brave intrepid explorers installed
> Topspin 5.0 yet?
>
> *This email originated outside the University. Check before clicking
> links or attachments.*
>
> Hi Sameer,
>
> I got it working.  There are some prompts and messages during the
> install that are very boring (I'm blaming them for the fact that I
> didn't read them carefully enough) that I should have paid more
> attention to.
>
> Once I figured that out, I was in.  And it imported all of my previous
> settings, especially all of my users and custom ICON settings
> seamlessly. It's working great.  Frankly it went more smoothly that most
> of the updates I've done over the past 15 years - and I've done every
> one of them.  I'm glutton for punishment.
>
> Setting up SpinPilot will take more time, but if you just want to update
> to 5.0 and keep using ICON, it's really rather painless.
>
> Ken Sharp-Knott
>
> Manager of Analytical Services and the NMR Facility
>
> Department of Chemistry
>
> Virginia Tech
>
> (540)267-6502 (Cell)
>
> (540)231-0885 (Office)
>
>

--
Dr. rer. nat. Matthias Brandl (he/him or they/them)

Technical University of Munich
TUM School of Natural Sciences - BNMRZ
Chair of Biomolecular NMR Spectroscopy (Prof. Sattler)

Lichtenbergstr. 4 (postal)
Ernst-Otto-Fischer-Str. 2 (physical)
85748 Garching b. Muenchen

Phone +49 89 289-52617

matthias.brandl_at_tum.de
https://www.sattlerlab.de/
https://www.bnmrz.org/



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3228): https://ammrl.groups.io/g/main/message/3228
Mute This Topic: https://groups.io/mt/117761947/7559972
Group Owner: main+owner_at_ammrl.groups.io
Unsubscribe: https://ammrl.groups.io/g/main/leave/12177410/7559972/1125599165/xyzzy [walt_at_hawaii.edu]
-=-=-=-=-=-=-=-=-=-=-=-


Received on Tue Feb 17 2026 - 03:54:34 MST