Re: [AMMRL] : CodeMeter on RockyLinux 9

From: Zhang, Weixing via groups.io <Weixing.Zhang=stjude.org_at_groups.io>
Date: Tue, 10 Feb 2026 15:17:53 +0000

Thank you, Clemens, and thank you to everyone else for your help.
CodeMeter is now installed, and TopSpin 5.0.0 is running correctly.

Best regards,

Weixing

> From: main_at_ammrl.groups.io On Behalf Of Anklin, Clemens via groups.io
> Sent: Tuesday, February 10, 2026 5:34 AM
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] : CodeMeter on RockyLinux 9



Hi Weixing,

Greg is correct you can download the latet client from wibu.com directly.
But this might not be sufficient. If after installation you do not have some
container number then you need to install a default container before activating
the license. I will send this to you directly.

You should be able to install this container by either loading it from the
control center with the File -> Import License function or from the command
line with "cmu -i -f<PATH_TO_FILE>"

Greg is also correct that you need to release the license from the old computer.
If you no longer have the ticket number send me the container number from the
old computer and we can find it for you. The container number is something
starting in 128-...

Best regards

Clemens


On 2/9/2026 3:02 PM, gregory.wylie via groups.io wrote:

Hi Weixing

I have not used Rocky, but codemeter can be downloaded separately.
I think this is correct?
https://www.wibu.com/us/products/codemeter/runtime.html


There are other issues if this is replacing an existing computer running codemeter
you will need to release the license from the old system and reinstall it on
the new build. To do this you need the container number. I had to get Bruker
to give me that on all my systems. This is the like 16 digit code you get on
Brukers website when you are setting up licenses. I never new we needed to
keep it until we started rebuilding. This may have changed but like 2 years
ago you had to do it this way.


Greg

Gregory P. Wylie, Ph.D
NMR Facility Manager
Texas A&M University
Department of Chemistry #3255
580 Ross St.
College Station, TX 77843-3255
gpwylie_at_tamu.edu
979.458.0705 (voice)
979.845.4719 (fax)
706.206.0007 (cell)
http://nmr.tamu.edu



-- =========================
==========================
=====================
Clemens Anklin Ph.D. Vice President
Bruker BioSpin NMR Applications & Training
15 Fortune Drive
NEW Phone: 978-313-5NMR(5667) Billerica MA 01821
bluesky: _at_canklin.bsky.social
web: http://www.bruker.com
e-mail: clemens.anklin_at_bruker.com
==========
==========================
==========================
===========


________________________________

Email Disclaimer: http://www.stjude.org/emaildisclaimer
Consultation Disclaimer: http://www.stjude.org/consultationdisclaimer


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3208): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3208__;!!PvDODwlR4mBZyAb0!U6UkvTPzFPWUgiLCwIQaRcpbsdwo0s=
iUjmWR80QvBuBvG18w43WGPuTX_Ks1Tzig5uAKoOCQh8wqB7NqM1u1UZ3h-Wjn$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/117725676=
/7559972__;!!PvDODwlR4mBZyAb0!U6UkvTPzFPWUgiLCwIQaRcpbsdwo0siUjmWR80QvBuBvG=
18w43WGPuTX_Ks1Tzig5uAKoOCQh8wqB7NqM1u1UULtqm06$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Feb 10 2026 - 07:18:52 MST

This archive was generated by hypermail 2.4.0 : Sun Mar 01 2026 - 13:47:31 MST