HI Igor
are tftpd and dhcpd running? Are all the required 32 bit compatibility packages installed?
https://repo-bbio.bruker.com/AlmaLinux/9/
Alma 9 is only tested with Topspin 3.7. I am aware, that Avance II requires
topspin 3.2. We can not guarantee that it works.
Regards
Clemens
On 2/6/2026 3:59 PM, D'Anciaes Almeida Silva, Igor via groups.io wrote:
Dear community,
I've being struggling all week to bring the workstation that controls a Avance II 300 k
to work with Alma9 but I'm getting stuck on an IPSO post code "b2". Apparently, IPSO
is getting an .100 IP address instead of the usual .98, and it's not liking it. Ideas?
PS: We already tried booting the workstation with the console down and turning only
the IPSO/AQC rack on (not possible to boot IPSO alone) and booting the workstation
with the console full on.
----------------
Dr. Igor d'Anciães Almeida Silva
Research Scientist - NMR Core
University of Missouri-Columbia
32A Chemistry Building
--
==========================
==========================
====================
Clemens Anklin Ph.D. Vice President
Bruker BioSpin NMR Applications & Training
15 Fortune Drive NEW Phone: 978-313-5NMR(5667)
Billerica MA 01821 bluesky: _at_canklin.bsky.social
web: http://www.bruker.com
e-mail: clemens.anklin_at_bruker.com
==========================
==========================
====================
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3199): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3199__;!!PvDODwlR4mBZyAb0!RnDtt9YMhH6HCcxAqSxBgWzjL6ktz5=
whW2xBkeiPlxOn0P60f6FyuHkIXY2XaLSEC7M5q8TWYZat2br8Tno50uMpbtiM$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/117680202=
/7559972__;!!PvDODwlR4mBZyAb0!RnDtt9YMhH6HCcxAqSxBgWzjL6ktz5whW2xBkeiPlxOn0=
P60f6FyuHkIXY2XaLSEC7M5q8TWYZat2br8Tno50tTjy0FU$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Sat Feb 07 2026 - 17:46:27 MST