Hello All,
Fairly sure this was discussed before but I wish to quickly share the information
of this. After many years the transfer lines that I had been using for my Bruker
800 and 900 pumped magnets had deteriorated. I was not really interested in what
the original company Messer or Bruker had as far as options or costs.
The solution seems easy at first but simply I was not sure if they wanted to do it.
I went to Cryofab. Was able to send in the line and they repaired a leak. Also,
at my request they cut the proprietary actuators replacing them with theirs.
For about $800US including shipping I have a essentially new line that I can pump in
the future with my own setup. Hope this information is useful.
Mike Delk
NMR Manager
University of Pittsburgh School of Medicine
Department of Structural Biology
412.370.0066--
-=-=-=-=-=-=-=-=-=-=-=-