Hi Igor
does zg with the same parameters work. Is it only WOBB? Can you please try
to change the parameters WBSW and WBST. This is the wobble sweep width and
wobble steps. For example set it to 4 MHz and 512 steps. This forces a
recreation of the wobble lists.
Clemens
On 2/27/2024 12:41 PM, D'Anciaes Almeida Silva, Igor wrote:
Dear Simon and Casey,
I just set the 119Sn parameters in the prosol table to the same values as 31P
but error remains. Anything else I could try?
Best,
----------------
Dr. Igor d'Anciães Almeida Silva
Research Scientist - NMR Core
University of Missouri-Columbia
32A Chemistry Building
________________________________
> From: main_at_ammrl.groups.io on behalf of Sham, Shing via groups.io
> Sent: Tuesday, February 27, 2024 11:25 AM
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] Error 152 in RX_1: CD SLOT 3 RACK1: invalid rwo number in loop table
Hi Igor,
Did you set-up 119Sn parameters in “prosol” table?
Good luck.
Simon
--
==========================
==========================
====================
Clemens Anklin Ph.D. Vice President
Bruker BioSpin NMR Applications & Training
15 Fortune Drive NEW Phone: 978-313-5NMR(5667)
Billerica MA 01821 twitter: _at_canklin
web: http://www.bruker.com
e-mail: clemens.anklin_at_bruker.com
==========================
==========================
====================
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1028): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1028__;!!PvDODwlR4mBZyAb0!RLVUL4eLgQ5zrnc-c7niK16ZqmAswL=
KdVhumm4hxFUwfeIbCB9uvYXES0kMTTn7wn2_XLQV8YfaHi-cwh3WI04C6YMd9$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/104606619=
/7559972__;!!PvDODwlR4mBZyAb0!RLVUL4eLgQ5zrnc-c7niK16ZqmAswLKdVhumm4hxFUwfe=
IbCB9uvYXES0kMTTn7wn2_XLQV8YfaHi-cwh3WI04wuWzZe$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Tue Feb 27 2024 - 10:50:28 MST