Re: [AMMRL] problems with pulsecal --> One way valves for helium exhaust

From: <brcherry_at_asu.edu>
Date: Wed, 21 Feb 2024 12:57:26 -0800

I've attached the last 24 hrs of data from the boil off on our non-pumped magnets. 
The three Bruker magnets have their standard one way valve with an adaptor to
KF-25.  The Varian's 400 and 400 WB both have the Z58140.   All of
the magnets have some level of oscillation, but no real differences between them.

If anyone is curious, we have integrated some AAlborg mass flow meters. (XFM07
and XFM17) to record this data.


-=-=-=-=-=-=-=-=-=-=-=-