Hi Rajan, HI Sameer
yes it is only the newest versions of MacOS that suffer from this issue. It
has to do with the linker. We have resolved the issue and it will be included
in the soon to be released Topspin 4.4.0
This is not a problem with topspin. It has to do with the fact that we can
not predict what changes Apple is introducing with new versions. Forward
compatibilty is a myth.
Topspin AU programs are not the only things that have problems.
Wait for Topspin 4.4.0 to appear and you can compile again.
Clemens
On 2/14/2024 2:09 PM, DrSpin wrote:
Hi Sameer
Here is my 2c. Although I do have Xcode in my MacBook installed, I don't
recall installing that first as a requirement for TS4.3.0 to run (it's been
a while; so forgive me if I am quoting this wrong). AU programs have been
compiling OK on my system.
My OS version is : Sonoma 14.1.2
Is there is an issue of compatibility with the later OS release ?
Rajan
On Wed, Feb 14, 2024 at 9:43 AM Sameer Al-Abdul-Wahid wrote:
Dear AMMRL,
When running TopSpin 4.3.0 on a Mac (OS Sonoma 14.3.1, M1 Pro) and compiling
an AU program (e.g., xiref), I get an error:
-macosx_version_min has been renamed to -macos_version_min
ld: unsupported mach-o filetype (only MH_OBJECT and MH_DYLIB can be linked)
in '/opt/topspin4.3.0/gnu/lib/libgcc_ext.10.5.dylib'
collect2: error: ld returned 1 exit status
I have XCode Command Line Tools installed on my computer, and tried a fresh
install of 4.3.0 just in case. Is this a problem other Mac users are having,
and does anyone know of a workaround?
Thanks!
Sameer
--
Sameer Al-Abdul-Wahid, Ph.D. | Manager
Nuclear Magnetic Resonance Centre | University of Guelph
519-824-4120 Ext. 58914 | s.wahid_at_uoguelph.ca<mailto:s.wahid_at_uoguelph.ca>
http://nmr.uoguelph.ca/
--
____________________________________
Rajan K Paranji, Ph.D. NMR Services Manager Department of Chemistry Room 65,
Bagley Hall University of Washington Seattle, WA 98195 phone : 206 685 2581
fax: 206 685 8665 email: rajanp_at_uw.edu<mailto:rajanp_at_uw.edu>
_____________________________________
--
==========================
==========================
====================
Clemens Anklin Ph.D. Vice President
Bruker BioSpin NMR Applications & Training
15 Fortune Drive NEW Phone: 978-313-5NMR(5667)
Billerica MA 01821 twitter: _at_canklin
web: http://www.bruker.com
e-mail: clemens.anklin_at_bruker.com
==========================
==========================
====================
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#983): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/983__;!!PvDODwlR4mBZyAb0!UJChfElnki3ZLeYngmppXQlnUEE7_eb-=
ilyc_HSxu_EYKAxZIuuCLih_WdqJwWNA73oT0liETT5O5PnRkJDiM1bY7fF7$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/104357463=
/7559972__;!!PvDODwlR4mBZyAb0!UJChfElnki3ZLeYngmppXQlnUEE7_eb-ilyc_HSxu_EYK=
AxZIuuCLih_WdqJwWNA73oT0liETT5O5PnRkJDiMyyYnmGs$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Wed Feb 14 2024 - 16:22:44 MST