Hi Lenny,

Later versions of the Bruker VT manual (Z33073) go into detail about operating
at lower temperatures – well worth a read.

The shim coils will operate quite happily down to sub-zero temperatures provided
you’re purging the shim coils with dry gas to prevent ice build-up.
The manual shows some very reassuring data with the probe set at -150 °C
and the shim coil at -52 °C without issue, for >2 days.

Thanks

Geoff

> From: main_at_ammrl.groups.io On Behalf Of Michael Groves via groups.io
> Sent: 17 December 2025 14:21
> To: main_at_ammrl.groups.io
> Subject: [External] Re: [AMMRL] Coil temp cooling too fast during LN₂ VT experiments


Hi Lenny,

It is my understanding that the main concern at low temperatures is ice build-up,
so if you're blowing dry N2 gas around the shims you should be safe even if
the temperature drops lower than you'd like. It's just at elevated temperatures
that the epoxy in the shim coil can soften and the shim coils will tend to move around.

On the newer Bruker spectrometers there's a ring that goes around the top of
the shim system that replaces the one on the older spectrometers--it's got
a port for a 6mm air line that is used to blow air down around the shim system.
I'm not sure what your setup looks like but you could install one of those if
you've got one--that might improve the efficiency of the shim gas.

One other consideration--that function is automatic on the newer Bruker systems
and it tends to use quite a bit of gas. I don't believe an AVII has a separate
valve for that so you've likely got it set up manually. I would recommend
cranking the flow up. You don't want it going full blast, but try increasing
the flow.

That's what I've got.

Cheers,
Mike

On Wednesday, December 17, 2025 at 04:19:34 AM MST, Lenny Haddad via groups.io wrote:

Dear all,

I’m running a low-temperature experiment on a Bruker Avance II with
 a BBO probe, using LN₂ VT cooling.

During cooldown, I’m purging/aerating the shim stack with dry N₂
(instead of air) to avoid ice formation. However, the temperature around
the shims drops very quickly and gets down to about 5 °C well before
the sample reaches the target temperature of −70 °C. I’m
a bit concerned about overcooling the shim stack and nearby bore components.

I was wondering if you have any recommendations or best practices for keeping
the shim stack above 0 °C while running low-temperature experiments.

Thank you,

Best regards,

Lenny

-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3068): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3068__;!!PvDODwlR4mBZyAb0!RST1LQQjUCzUHLMNHOak_DG171QJwN=
f2ktdpXqM_GZa6ShWTfkey0oq19j_cCwEQh3sM1eAVwjIbQpkk8VmSBT6kXtr6$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/116824723=
/7559972__;!!PvDODwlR4mBZyAb0!RST1LQQjUCzUHLMNHOak_DG171QJwNf2ktdpXqM_GZa6S=
hWTfkey0oq19j_cCwEQh3sM1eAVwjIbQpkk8VmSBa6ZTIYh$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Dec 17 2025 - 06:48:18 MST

This archive was generated by hypermail 2.4.0 : Thu Jan 01 2026 - 15:24:06 MST