Re: [AMMRL] Sample insertion issues

From: Spin-Doc <bert.heise_at_spin-doc.net>
Date: Wed, 17 Dec 2025 10:39:41 +0100

Hi,

in case oyu system having a room-temperature probe, try to remove it and try
to take pictures down its top bore tube (where the sample is inserted into)
with a light and/or flash in order to spot whether the inner glass insert
tube (of very thin wall thickness, only about 1/2 mm) shows signs of being
cracked: that might be another source of samples randomly not inserting
all the way… (let's hope not - otherwise the probe needs to be
repaired)

Cheers
Bert

> Am 16.12.2025 um 22:43 schrieb Michael Groves via groups.io:
>
> Hi Eugenio,
>
> Normally the 250lph gasflow is plenty low enough to allow the sample to
> seat. There are a couple of things that come to mind:
>0
> Is the sample really descending or is it getting caught at the interface
> between the sample transport and the shim system? If that gets a little
> bit out of alignment you can start getting things caught there. If it
> gets past the interface and really is dropping down into the magnet then
> it's likely not an alignment issue.
>
> Do you have a cryoprobe? Is the emergency VT gas connected? If so, you
> might try disconnecting that from the input to the probe. (if it's there,
> there's a one-way valve and then an adapter and then the tubing to the
> console. Just pull the tubing out--leave the one-way valve installed.
> If the emergency VT gas is too high then it can keep the sample from
> seating properly.
>
> I believe there's a timeout that you can adjust in the BSMS service webpage
> that will increase the time it waits for the sample to seat--you could try
> increasing that to see if a longer delay (60s? Longer?)
>
> Is the VT gasflow really being decreased? If you open edte and watch the
> gasflow, and then insert a sample, does it really drop to 250lph?
>
> And, there have been cases where people have overheated their spinners
> and they go out of round. One possibility would be to take each spinner
> that fails out of the rotation. That way you can see if the failures
> really are random or if there is some group of spinners that just
> doesn't want to insert correctly.
>
> That's what I've got.
> Cheers,
> Mike
>
> On Tuesday, December 16, 2025 at 01:53:40 PM MST, Eugenio Alvarado via groups.io wrote:
>
> Dear all,
>
> Recently, our Avance Neo 500 with a SampleCase is having more frequent
> random issues with about 10% of samples that don't seat completely
> inside the magnet. The samples lift and descend properly in the magnet
> but sometimes they just don't settle down properly, even after a delay
> of 30 or 45 seconds, causing IconNMR to eject and try inserting the
> sample several times until it gives up. When that happens, not only
> that sample fails, but also the next one in the queue, probably because
> its start time has timed out after the previous attempts. The gas flow
> pressure in the console is set at 4.5 bar and I just changed the "VT gas
> flow limitation" in the BSMS Web Services interface from 700 to 250 with
> just a little improvement. The VT gas flow in standby is 40 l/h and
> for operation it is 400.
>
> Can anyone please suggest any other parameter to adjust to make sample
> insertions more reliable? Or could it be a problem with a valve?
>
> Thanks in advance,
> Eugenio
>
> --
> Eugenio Alvarado, Ph.D.
> NMR Facility Director
> Chemistry Department, room 3500. University of Michigan
> 930 N. University Ave., Ann Arbor, MI 48109-1055.
> Voice: (734)-764-9924
>
>


Magnetic resonance equipment, service, training & more
Selling & buying used NMR equipment (all vendors)
Dr. rer. nat. Bert Heise
Owner, Applications Scientist

Spin-Doc Bert Heise
Kiefernweg 13
58239 Schwerte
Germany
Mobile: +49 171 5135315
Email: bert.heise_at_spin-doc.net
Web: http://www.spin-doc.net

Representing:
Stelar Fast Field Cycling NMR Relaxometers (http://www.stelar.it)
MestreLab NMR Software (http://www.mestrelab.com)
Barthel HF Amplifiers (http://www.barthel-hf.de)


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3065): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3065__;!!PvDODwlR4mBZyAb0!XK_Vtg_0LQYCsDa3226EYpMzX8ItmY=
xvQIkPQL_aJ1EqCiI3pvVsIFKGsyydXwwUYMxWKx2tYPeTli58v4SQ0w$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/116816512=
/7559972__;!!PvDODwlR4mBZyAb0!XK_Vtg_0LQYCsDa3226EYpMzX8ItmYxvQIkPQL_aJ1EqC=
iI3pvVsIFKGsyydXwwUYMxWKx2tYPeTli7ZEDq-6g$
Group Owner: main+owner_at_ammrl.groups.io
Unsubscribe: https://urldefense.com/v3/__https://ammrl.groups.io/g/main/lea=
ve/12177410/7559972/1125599165/xyzzy__;!!PvDODwlR4mBZyAb0!XK_Vtg_0LQYCsDa32=
26EYpMzX8ItmYxvQIkPQL_aJ1EqCiI3pvVsIFKGsyydXwwUYMxWKx2tYPeTli7VPDXt9g$ [wa=
lt_at_hawaii.edu]
-=-=-=-=-=-=-=-=-=-=-=-



Received on Wed Dec 17 2025 - 03:19:14 MST

This archive was generated by hypermail 2.4.0 : Thu Jan 01 2026 - 15:24:06 MST