Re: [AMMRL] Sample insertion issues

From: Charles Fry via groups.io <cfry03=esf.edu_at_groups.io>
Date: Tue, 16 Dec 2025 21:30:58 +0000

Eugenio,

We were having exactly the same problem here recently. It started when I
changed us from a BBFO+ probe to a Prodigy BBO probe about two months ago.
I have been using the Bruker recommended (PICS) value of 535 for the VT gas
flow, but on reducing it to 450 a couple weeks ago the problem has (so far)
disappeared. On another spectrometer at UW-Madison the prescribed VT flow
caused samples to lift to unlocked positions. I'm not confident about this
fix, and reducing the VT flow might give some other concerns (esp on a
cryoprobe). Be grt to hear other suggestions.

Charlie

________________________________
> From: main_at_ammrl.groups.io on behalf of Eugenio Alvarado via groups.io
> Sent: Tuesday, December 16, 2025 3:21 PM
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: [AMMRL] Sample insertion issues

Dear all,

Recently, our Avance Neo 500 with a SampleCase is having more frequent
random issues with about 10% of samples that don't seat completely
inside the magnet. The samples lift and descend properly in the magnet
but sometimes they just don't settle down properly, even after a delay
of 30 or 45 seconds, causing IconNMR to eject and try inserting the
sample several times until it gives up. When that happens, not only
that sample fails, but also the next one in the queue, probably because
its start time has timed out after the previous attempts. The gas flow
pressure in the console is set at 4.5 bar and I just changed the "VT gas
flow limitation" in the BSMS Web Services interface from 700 to 250 with
just a little improvement. The VT gas flow in standby is 40 l/h and
for operation it is 400.

Can anyone please suggest any other parameter to adjust to make sample
insertions more reliable? Or could it be a problem with a valve?

Thanks in advance,
Eugenio

--
Eugenio Alvarado, Ph.D.
NMR Facility Director
Chemistry Department, room 3500. University of Michigan
930 N. University Ave., Ann Arbor, MI 48109-1055.
Voice: (734)-764-9924


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3062): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3062__;!!PvDODwlR4mBZyAb0!TY-FqlH7V3AVX6dfeyCCiV0HJtXVyV=
zObieIBrxq55ub3_oy3W8lHAxXB78cyyW4jpFuXm2ZqQfDFaLdFGwLQwAVDA$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/116816512=
/7559972__;!!PvDODwlR4mBZyAb0!TY-FqlH7V3AVX6dfeyCCiV0HJtXVyVzObieIBrxq55ub3=
_oy3W8lHAxXB78cyyW4jpFuXm2ZqQfDFaLdFGw2qijLtA$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Dec 16 2025 - 15:21:40 MST

This archive was generated by hypermail 2.4.0 : Thu Jan 01 2026 - 15:24:06 MST