Re: [AMMRL] Replacing an AVANCE II PC in 2025

From: Laird Trimble via groups.io <laird.trimble=trimbleanalyticalservices.com_at_groups.io>
Date: Wed, 10 Dec 2025 16:34:35 -0500

Hi Sameer,

I have Avance II systems running in a VM in VirtualBox. One is running CentOS 5
as the guest OS and TopSpin 2.1. The other is running CentOS 7 as the guest OS
and TopSpin 3.2. In the former case Ubuntu 24.04LTS is the host and Win11 is
the host for the latter. Both systems are running well and have had few problems.
In order to keep corporate IT departments out of my NMR spectrometers, I have
always preferred using Linux as the host but your requirements may be different.

One problem that my users do run up against occasionally is that they get confused
between the host and guest OS’s and TopSpin winds up being in a small or
minimize window which can lead to confusion. You can go a long way to hide the
host OS but you know users, if they find a button, they gotta click on it.

Other than that, this has been a reliable solution in my hands.

Cheers,

Laird

Laird A. Trimble, Ph.D.
President, Trimble Analytical Services Ltd.
laird.trimble_at_trimbleanalyticalservices.com
1-613-286-2799



> On Dec 10, 2025, at 3:41 PM, Sameer Al-Abdul-Wahid via groups.io wrote:
>
> Hello AMMRL,
>
> Our AVANCE II computer died and I’m trying to replace it.
>
> At the suggestions of others, I’ve started down the route of setting
> up VirtualBox on our Windows 11 machine, so I can create a CentOS 5 VM to
> run TopSpin 2.1. It’s a “two steps forward, one step back”
> process - for example, I can install CentOS 5, but doing an update of the
> (non-Bruker) packages requires a workaround.
>
> Before I go too far down this path, I thought I’d ask the collective
> wisdom - what have others done here? Are there newer Linux OSs that could
> work here? Using Windows seems to rely on needing an NFS server that I
> cannot seem to find.
>
> Our system could in theory support TopSpin 3.2 so I’d be open an
> option that required me to use 3.2 as well.
>
> Thanks for your time,
> Sameer
>
> --
> Sameer Al-Abdul-Wahid, Ph.D.
> Manager, Nuclear Magnetic Resonance Centre <http://nmr.uoguelph.ca/>
> University of Guelph
> 519-824-4120 x58914 / s.wahid_at_uoguelph.ca
>



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3039): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3039__;!!PvDODwlR4mBZyAb0!TMHiRtlDNQF1PNZxrnjVtNXtMg_IRp=
MfukQqIJBGuUBmlFwi8tCX-avUU_tiWnFkV1wmx7CuYt7Z95C_ZEVkyZN0j0un$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/116718929=
/7559972__;!!PvDODwlR4mBZyAb0!TMHiRtlDNQF1PNZxrnjVtNXtMg_IRpMfukQqIJBGuUBml=
Fwi8tCX-avUU_tiWnFkV1wmx7CuYt7Z95C_ZEVkySwJqq2u$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-



Received on Wed Dec 10 2025 - 19:05:16 MST

This archive was generated by hypermail 2.4.0 : Thu Jan 01 2026 - 15:24:06 MST