Date: Wed, 11 Dec 2024 19:42:40 +0100
Hi Jerry,

This largely depends on which preamp exactly you want to replace.
Do you have a “correct frequency” preamp of the same type available (even if it’s broken)?

- A X-BB preamp with exchangeable external filter unit should work at any frequency
- a highband/1H preamp will be very difficult because of internal filters
- the newer the preamp is (HPPR/HPPR2/3) the less likely it will work because the filters got more and more integrated into the board itself instead of being possibly swappable filters
- lock preamps might be doable but the filter units in the side need to be exchanged
- newer X band filters tend to have a built-in deuterium reject filter which may block desired frequencies (and you need to add an external lock reject filter for your lock frequency)

In German, we call this “Jein” (“yeahno”)… ;-)
Bert

-> Sent via mobile <-

Spin-Doc 
Magnetic resonance equipment, service, training & more
Dr. rer. nat. Bert Heise
Owner, applications scientist
Kiefernweg 13
58239 Schwerte
Germany
Mail: bert.heise_at_spin-doc.net
Phone: +49-171-5135315

Representing MestreLab, Stelar & Barthel HF



Am 11.12.2024 um 19:11 schrieb Jerry Hu via groups.io :



Dear AMMRL,

 

Has anybody tried replacing a HPPR preamplifier with one for a different frequency? For example, replace a 500MHz preamplifier with a 400MHz or 600MHz one. I appreciate your inputs.

 

Cheers,
Jerry
-----------------------------------------------
Jerry Hu, Ph.D.
Technical Director, Project Scientist
Materials Research Lab, UCSB
Santa Barbara, CA 93106
Jerry.hu_at_ucsb.edu
(805)893-7914 (office)
(805)893-7940 (lab)

 

_._,_._,_

Groups.io Links:

You receive all messages sent to this group.

View/Reply Onl= ine (#1937) | Reply = to Group | Reply to Sender | Mute This Topic | New Topic
Your Subscription | Contact Group Owner |

_._,_._,_
--Apple-Mail-D3FE78A4-F1DD-4448-BCB1-244A00099B40-- Received on Wed Dec 11 2024 - 10:43:01 MST

This archive was generated by hypermail 2.4.0 : Sun Jan 05 2025 - 13:37:38 MST