Re: [AMMRL] BACS-120 pincher

From: Craig Grimmer <craig.grimmer_at_gmail.com>
Date: Sun, 10 Dec 2023 08:45:48 +0200

Hello All

When the rubber grip on the end of the fingers wears away, a little *thin*
heat-shrink tubing can be used as a substitute.

Kind regards,
Craig.


On Mon, 4 Dec 2023 at 21:01, dennisonpr <dennison_at_uci.edu> wrote:

> I believe that o-rings were used originally, mine had one too. The spring
> was an later improvement.
>
> Phil.
>
> On 20231204 09:57, Perry Pellechia wrote:
>
> I never knew this was originally a spring. I found an o-ring during my
> first repair. Glad you were able to fix the issue.
>
> Perry.
>
> On Dec 4, 2023, at 10:19 AM, Justin Pontius via groups.io
> <jpontius=nd.edu_at_groups.io> <jpontius=nd.edu_at_groups.io> wrote:
>
> Wow! Many thank you's Perry!
>
> I took it apart and indeed a broken spring popped out. I actually found a
> replacement round spring in the blue box. All fixed!
> https://photos.app.goo.gl/XKn8HhCPBA9WYjim6
>
> Thanks again!
>
> --Justin
>
>
> <IMG_1362.HEIC>
>
>
>
> --
> Perry.
> --
>
> Perry J. Pellechia
> Director of NMR Services
> Research Professor
> University of South Carolina
> Department of Chemistry and Biochemistry
>
>
>
>
>


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#861): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/861__;!!PvDODwlR4mBZyAb0!RGFl1k6g4fafD1FGukn8oaN4SDgHTgXh=
JoHmikEztOK41b3VIVkJ6N2cnG1Cxnd1guvZr-o5zyt_zkWnKNvArg$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/102970731=
/7559972__;!!PvDODwlR4mBZyAb0!RGFl1k6g4fafD1FGukn8oaN4SDgHTgXhJoHmikEztOK41=
b3VIVkJ6N2cnG1Cxnd1guvZr-o5zyt_zkUEqq0f9w$
Group Owner: main+owner_at_ammrl.groups.io




Received on Sat Dec 09 2023 - 22:46:05 MST

This archive was generated by hypermail 2.4.0 : Mon Jan 01 2024 - 19:06:37 MST