Re: [AMMRL] BACS-120 pincher

From: DrSpin <rajanp_at_uw.edu>
Date: Mon, 4 Dec 2023 07:53:30 -0800

Hi Justin and Perry
   Thank you for this thread, as I am also contending with an aging BACS120
and sundy. I recently had to replace one of the five fingers as it was
getting 'weak' i.e. it won't grab onto the tube and the tube will be skewed
from its axial position. When the vertical arm came down the tube will
descend quite off center from the magnet and this led to tube breakage a
few times.

   In my case, it turned out that the faulty finger had a broken dowel pin
and so it was not moving the full range. Also, there is a rubber/polymer
lining on the top of the finger that provides a bit of tactile touch so
that the steel finger doesn't crush the tube. Due to wear and tear, this
lining had completely gone in my case.

I am glad to hear about your ending and also good to know that the spring
that controls the motion of all the five pinchers/fingers can break too.

Best Regards
Rajan

On Mon, Dec 4, 2023 at 7:42 AM Justin Pontius via groups.io <jpontiusnd.edu_at_groups.io> wrote:

> Wow! Many thank you's Perry! I took it apart and indeed a broken spring
> popped out. I actually found a replacement round spring in the blue box.
> All fixed! https: //photos. app. goo. gl/XKn8HhCPBA9WYjim6 Thanks again!
> --Justin
>
> Wow! Many thank you's Perry!
>
> I took it apart and indeed a broken spring popped out. I actually found a
> replacement round spring in the blue box. All fixed!
> https://photos.app.goo.gl/XKn8HhCPBA9WYjim6
>
> Thanks again!
>
> --Justin
>
>
>
>
>

--

____________________________________

Rajan K Paranji, Ph.D.NMR Services Manager
*Department of Chemistry**Room 65, Bagley Hall*
University of WashingtonSeattle, WA 98195
phone : 206 685 2581 fax: 206 685 8665email: rajanp_at_uw.edu
____________________________________________________________________


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#811): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/811__;!!PvDODwlR4mBZyAb0!SypCSdbrIlJXGTY_F8T180wo1Kj0waLi=
h-RZTIBPDTm-gvRIwELbyjN633GLN5p9sLR-DmBv2sM9cw$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/102970731=
/7559972__;!!PvDODwlR4mBZyAb0!SypCSdbrIlJXGTY_F8T180wo1Kj0waLih-RZTIBPDTm-g=
vRIwELbyjN633GLN5p9sLR-DmC0z3CHPQ$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon Dec 04 2023 - 07:54:13 MST

This archive was generated by hypermail 2.4.0 : Mon Jan 01 2024 - 19:06:37 MST