Setup on Unity 400, Incorrect reffrq

From: Russell, Vernon <Vernon.Russell_at_omnova.com>
Date: Wed, 1 Dec 2004 13:07:47 -0500

When running an experimental setup from the VNMR menu, or running in
automation, our Unity 400 changes reffrq to a value based on lockfreq=1.144.
This gives reffrq=7.4524194 MHz for 1H acquisition, and a ridiculous
chemical shift scale. Loading the H1.par parameters directly into an exp and
acquiring data works fine and doesn't change reffrq (reffrq remains at
399.942). For Unity systems lockfreq=1.144 is a correct setting.

 

Does anyone know how to correct this problem?

 