Summary: Maintenance of Transfer Tube Vacuum

From: Edward E. Waali <waali_at_selway.umt.edu>
Date: Fri, 20 Dec 2002 15:55:31 -0700

NMR Folks,

         Thanks, for the nearly 40 responses received in the first day.

         About 42 % routinely re-evacuate the helium transfer line every 6
months to 2 years. A few do it before each fill.

         About 34 % re-evacuate only after cold spots or frost are
noticed. Still, 2 to 12 years passed before the cold spots developed.

         About 24 % have never re-evacuated the line over periods of 3 to
17 years.

         Of those who re-evacuate, about 1/2 use a standard rotary pump,
letting the liquid helium cryopump the remaining air. The other 1/2 use a
turbo or diffusion pump. Practically everyone does it locally even if they
send it to another department on their campus. Most use a liquid nitrogen
trap between the pump and line.

         One said that the line becomes more flexible when it has lost the
vacuum.

Our solution: We will just watch for the cold spots. Several of you
pointed out the fitting with the built-in hex-adjuster which has always
been sitting in the Oxford magnet kit. We will use that, if necessary, to
avoid a Swagelock valve. Thanks, again!