Hi all,
This is a new problem that I have not seen reported before. The
spectrometer is a Vnmrs 600 with a OneProbe and OVJ 3.1A. When taking
proton spectra with Autogain on, vnmrj sets the gain much higher than
needed without triggering a receiver overflow warning. The resulting
spectra show the typical distortions of a receiver overflow on the
baseline. Reducing the gain by 10-18 units yields a perfect spectrum.
The autogain setting for my standard sample used to be around 30 but now
vnmrj sets it to 44. The pulse width and power are properly calibrated
and haven't changed in the last year or so. I would greatly appreciate
any idea of what to try.
Thanks,
Eugenio
--
Eugenio Alvarado, Ph.D.
NMR Facility Director
Chemistry Department, room 3500. University of Michigan
930 N. University Ave., Ann Arbor, MI 48109-1055.
Voice: (734)-764-9924
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3729): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3729__;!!PvDODwlR4mBZyAb0!SohQu6IjhgaXL3cnno1TB8zYm2_Hta=
vViJ7xXVlzomV-EMyONW8s1q5nzbd1uy1zvtT_2hVMZR7GdK3jP_5tJJ_pKHI$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121017517=
/7559972__;!!PvDODwlR4mBZyAb0!SohQu6IjhgaXL3cnno1TB8zYm2_HtavViJ7xXVlzomV-E=
MyONW8s1q5nzbd1uy1zvtT_2hVMZR7GdK3jP_5tL6Pcx2M$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Mon Aug 31 2026 - 10:08:56 MST