Rick
if it ain't broke don't fix it. If a magnet shows no significant change in
nitrogen and helium consumption it is good.
In order to reseal you have to de-energize and warm up. Nobody will guarantee
that it comes back up after this procedure. The chances are good but never 100%.
The procedure also costs a lot of helium.
With best regards
Clemens
On 8/28/2026 3:14 PM, Richard Fitch via groups.io wrote:
Dear Spinlanders,
I have a quandary on magnets and lifetimes. I have heard many times from the
group on the excellent lifetime for magnets but am curious as to the experience
of the group with refurbishing/resealing older magnets. We are batting around
the idea and it would be useful to know everyone’s successes and failures
with such things. I’m happy to post an anonymized summary if there
is sufficient interest.
All the best,
Rick
Richard W. Fitch
Professor, Department of Chemistry and Physics
Councilor, Wabash Valley Section, American Chemical Society
Indiana State University
600 Chestnut Street
Science Building, Rm. S35E
Terre Haute, IN 47809
Phone: (812) 237-2244
Fax: (812) 237-2232
“Do an apparently simple experiment that gives you an important bit
of information...
Julius Axelrod
The love of money is the root of some evil. Fear and insecurity are the root
of most of it.
--
==========================
==========================
====================
Clemens Anklin Ph.D. Vice President
Bruker BioSpin NMR Applications & Training
15 Fortune Drive NEW Phone: 978-313-5NMR(5667)
Billerica MA 01821 bluesky: _at_canklin.bsky.social
web: http://www.bruker.com
e-mail: clemens.anklin_at_bruker.com
==========================
==========================
====================
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3724): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3724__;!!PvDODwlR4mBZyAb0!Uk9BfrsPotIjdTi93QitilrFAKcX5A=
ChfvMpALRzSPcubnXHgW4dr6iAf9eXdbXx-82qNqmPFXF7WVGJ6EJs-QjNhKFI$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120977275=
/7559972__;!!PvDODwlR4mBZyAb0!Uk9BfrsPotIjdTi93QitilrFAKcX5AChfvMpALRzSPcub=
nXHgW4dr6iAf9eXdbXx-82qNqmPFXF7WVGJ6EJs-QwP5M1G$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Fri Aug 28 2026 - 13:01:28 MST