Hi Casey,
If your system is similar to all of our units, it is just a screw through
plastic. I assume you mean that the plastic is stripped. If that is the
case, I would cut new threads for a slightly larger screw.
Shawn
________________________________
> From: main_at_ammrl.groups.io on behalf of Casey Simons via groups.io
> Sent: Thursday, August 27, 2026 12:52 PM
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: [AMMRL] Varian ProTune Flexible Shaft
Hey all,
We have a ProTune Flexible Shaft that the screw will no longer tighten to
the dive shaft of the protune unit. Any chance does anyone have one up for
grabs before heading to the Agilent store or a good fix?
C
Casey Simons
NMR Director
CAMCOR Nanofabrication and Imaging Facility