Dear colleagues,
We have recently encountered an unusual issue with TopShim. The 'topshim 3d'
command has suddenly stopped working and now returns the following error:
getpar failed on /opt/topspin4.4.1/prog/curdir/nmrsu/topshimData/3D_acquisition_data/1/acqu2s
Parameter(s) not found in file: FnMODE
Of course, the 'acqu2s' file does contain this parameter, exactly as it used to
be when it was working before, and on other systems:
##$FnMODE= 0
The 'topshim 1d' command continues to work correctly, and normal 2D and 3D
experiments run without any issues. Also, there are no filesystem errors,
no updates were installed, etc.
At first, we were able to temporarily resolve the problem by rebooting both
the spectrometer and the workstation, followed by 'cf' and 'expinstall'. After
this, 'topshim 3d' would run successfully once, but any subsequent attempt
produced the same error again. After several attempts, even this workaround
stopped working.
Avance Neo
TopSpin version 4.4.1 running on CentOS 7
We would greatly appreciate any advice or suggestions
Thanks,
Andrei
-=-=-=-=-=-=-=-=-=-=-=-