Due to a recent magnet loss we found ourselves with an extra Avance III-HD
console, BBO probe, BST and shim stack. As “luck” would have
it, another department with an old failing Varian had a magnet of the same
strength. The starting magnet was an Oxford AS600 and the destination magnet
an unshielded Oxford compact flat bottom 600. Both with 51 mm bores and
having the centerline indicated as being the same distance from the bottom
in their respective manuals. The distance to the top on the unshielded magnet
differed by 9 mm so we made slight adjustments in the collar on the BST to
account for this. We configured (cf’d) the console on its new magnet
and entered the new field value which we had taken from the Varian sfrq
parameter. We’re searching for a lock signal but see none. So I
wonder if we either have the probe in the wrong place, sfrq doesn’t
equal sfo1 or a combination of both.
Can anyone recommend a systematic method for adjusting the probe height and
searching for a lock signal?
How far away from the sweet spot can the probe be but still show a signal?
Thanks,
John
John Grimes | NMR Facility Manager
Complex Carbohydrate Research Center, University of Georgia
315 Riverbend Road
Athens, GA 30602-4712
jgrimesjr_at_uga.edu ( jgrimesjr_at_uga.edu )
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3686): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3686__;!!PvDODwlR4mBZyAb0!SxKX6FLpfZL5EHVwk0rLks-k_e5-SB=
WyLoH7b3aSRdrJbIxenerzmhJlMP-o1QonK7KmvmcPXn4_WYqG-GKfT2pWSk4u$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120718739=
/7559972__;!!PvDODwlR4mBZyAb0!SxKX6FLpfZL5EHVwk0rLks-k_e5-SBWyLoH7b3aSRdrJb=
IxenerzmhJlMP-o1QonK7KmvmcPXn4_WYqG-GKfT8U931Cf$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Wed Aug 12 2026 - 07:50:01 MST