Re: [AMMRL] Spinners from Cortec.net?

From: Gerd Gemmecker via groups.io <Gerd.Gemmecker=tum.de_at_groups.io>
Date: Wed, 5 Aug 2026 10:38:32 +0200

I have a suspicion ... although the picture is then wrong, this might be for
the "special ultra-high field spinners" REQUIRED for GHz magnets,
that's why it explicitly says "modified standard spinner".

We have them on our 1200: they have (clearly visible!) little titanium rods
integrated in them to compensate for the Meissner-Ochsenfeld effect which lets
diamagnetic things (live frogs, K-TEL spinners etc.) "float" on very strong
magnetic fields. So "normal" K-TEL spinners will take forever to settle all the way down
in such a strong magnet. Understandably, these very special spinners would be
much more expensive than the normal ones (although the picture here definitvely
shows a normal one WITHOUT the titanium inserts -  I guess it was just copied
from the nromal spinners).

The "normal" spinners can be found here:
https://store.bruker.com/en-DE/products/spinner-standardbore-shimsystems?material=kel-f&size=5*2520mm
and they seem to be  one order of magnitude cheaper   ;-)

Ciao,

gg

--
PD Dr. Gerd Gemmecker
Bayerisches NMR-Zentrum
NAT School, TU München
Lichtenbergstr. 4
D-85747 Garching
Germany

Raum/Room 2108 / BNMRZ, E.O.Fischer-Str. 2
Tel. +49 (89) 289-52604
Fax +49 (89) 289-52669
e-mail: Gerd.Gemmecker_at_tum.de
Internet: http://www.gemmecker.de



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3642): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3642__;!!PvDODwlR4mBZyAb0!VygPiRN8pohC-t5MXDQhBSF14pKeRp=
NLkhs40eSxL8jOFYaKIlBrTYkIhVlo1qV5NFfD2Ehsq1awD1Xo7VelobqaMJ3c$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120586211=
/7559972__;!!PvDODwlR4mBZyAb0!VygPiRN8pohC-t5MXDQhBSF14pKeRpNLkhs40eSxL8jOF=
YaKIlBrTYkIhVlo1qV5NFfD2Ehsq1awD1Xo7Velodv7-ePK$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-


Received on Wed Aug 05 2026 - 02:43:11 MST

This archive was generated by hypermail 2.4.0 : Tue Sep 01 2026 - 13:53:27 MST