Re: [AMMRL] Spinners from Cortec.net?

From: Gary Thompson via groups.io <G.S.Thompson=kent.ac.uk_at_groups.io>
Date: Wed, 5 Aug 2026 14:56:07 +0000

The Kel-F spinners are often required for cryoprobes as they have a higher =
gas flow rate to avoid temperature gradient and th higher mass spinners prevent
‘wobbling’ and artefacts

You can also get mass multiplier rings to increase the weigh of blue spinners
to the same as Ken_f for $70 ish a pop

regards
gary

Dr Gary S Thompson NMR Facility Manager
CCPN CoI & Working Group Member
Wellcome Trust Biomolecular NMR Facility
School of Natural Sciences
University of Kent, Canterbury, Kent, England, CT2 7NZ

☎:01227 82 7117
✉️: g.s.thompson_at_kent.ac.uk
orchid: orcid.org/0000-0001-9399-7636




On 5 Aug 2026, at 15:41, Olson, Dean via groups.io wrote:

Group,

Bear in mind that the blue spinner has a mass of about 14.8 g. The Kel-F spinner
is about 21.6 g. I think we all know this, but these would not be expected
to be interchangeable.

Dean

***********************************
Dean L. Olson, Ph.D.
Director, NMR & EPR Labs
146 Roger Adams Lab, Box 16
School of Chemical Sciences
U. of Illinois at Urbana-Champaign
600 S. Mathews Avenue
Urbana, IL 61801 USA
Lab: 217-244-0564; Cell: 217-722-9432
dolson_at_illinois.edu
http://scs.illinois.edu/nmr
http://scs.illinois.edu/epr/
***********************************

London and South East University Group, a charity and company limited by gu=
arantee registered in England (company number 00986729). Registered Office:=
ld Royal Naval College, Park Row, Greenwich, London SE10 9LS


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3652): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3652__;!!PvDODwlR4mBZyAb0!QY0U4gLgExMsKn1W10OMceGWTS0_rB=
tEsYKiImEOQcwjzHavMMFC4taQ3QYBoZfL4fsPL6HG9E5Gttkd5g9906fVIjgF$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120586211=
/7559972__;!!PvDODwlR4mBZyAb0!QY0U4gLgExMsKn1W10OMceGWTS0_rBtEsYKiImEOQcwjz=
HavMMFC4taQ3QYBoZfL4fsPL6HG9E5Gttkd5g990xtKqCqo$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Aug 05 2026 - 07:59:29 MST

This archive was generated by hypermail 2.4.0 : Tue Sep 01 2026 - 13:53:27 MST