Re: [AMMRL] Spinners from Cortec.net?

From: Spin-Doc <bert.heise_at_spin-doc.net>
Date: Tue, 4 Aug 2026 21:28:47 +0200
Hi Anne,

just some 0.02 EUR (partially nitpicking…):
- it’s cortecnet.com (not cortec.net, easy to confuse)

- both POM and Kel-F are essentially „room-temperature“ spinners (Kel-F going 20
degrees lower & higher than POM) whereas only PEEK and ceramics offer the full -150…200 C VT range.
I myself used some Kel-F solids MAS turbines long time ago and they definitely started softening in the 50-60 deg C range…

- are you sure you compare the correct prices (I can’t believe Bruker sells RT-spinners for $2500 apiece)?

Cheers
Bert

-> Sent via mobile <-

Spin-Doc
Magnetic resonance equipment, service, training & more
Dr. rer. nat. Bert Heise
Owner, applications scientist
Kiefernweg 13
58239 Schwerte
Germany
Mail: bert.heise_at_spin-doc.net
Phone: +49-171-5135315

Representing MestreLab software, Stelar relaxometers & Barthel HF amplifiers



Am 04.08.2026 um 15:34 schrieb Vernon M Russell via groups.io <vrussell=wlgore.com_at_groups.io>:

I have a set of tan 5mm PEEK VT spinners from www.rototec-spintec.com. The temperature range is -150C to +200C.  They have worked well over many years and are much lighter than the heavy ceramic VT spinners and it is much easier to insert the a NMR tube.

The part number appears to be B-PEEK-5 and cost is 487.19 Euro. 


Vernon Russell

vrussell_at_wlgore.com
 
W. L. Gore & Associates, Inc.
Together, improving life

gore.com

 


_._,_._,_

Groups.io Links:

You receive all messages sent to this group.

View/Reply Onl= ine (#3639) | Reply to Group | = Reply to Sender | Mute This Topic | New Topic
Your Subscription | Contact Group Owner |

_._,_._,_
--Apple-Mail-FC1B636C-CDD0-400A-8002-B095031B6608-- Received on Tue Aug 04 2026 - 12:51:23 MST

This archive was generated by hypermail 2.4.0 : Tue Sep 01 2026 - 13:53:27 MST