Re: [AMMRL] OpenVnmrJ - problems with linking to NMR console

From: Shawn Wagner via groups.io <shawn.wagner=usc.edu_at_groups.io>
Date: Wed, 5 Aug 2026 21:26:17 +0000

Thanks Bert,

The link is great! I had only used the prepared builds. You are probably right
about compatibility. We only use our system to exclusively run OpenVnmr J and
software we write to manage user accounts and data transfer to our server.

For us, creating a /etc/udev/rules.d/70-persistent-net.rule file that assigns
the network card name at the startup has been very reliable. This way, we
always know which card belongs to the spectrometer. I looked up what you
meant by boot delays and found that the better systemd method for us should
be to use the /etc/systemd/network/10-custom-ethernet.link file. I like that
I know the spectrometer is eth1 and eth0 is my network connection.

Shawn
________________________________
> From: main_at_ammrl.groups.io on behalf of Spin-Doc
> Sent: Wednesday, August 5, 2026 1:20 PM
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: Re: [AMMRL] OpenVnmrJ - problems with linking to NMR console

Hi Shawn, I would rather advise against doing this (renaming the network ca=
rds in the old-fahioned eth0/eth1 way) unless REALLY necessary, like in ord=
er to make older Topspin 3. 2 cooperate (this requires eth0/1) with=
 newer Linux like Alma 8. 

Hi Shawn,

I would rather advise against doing this (renaming the network cards in the
old-fahioned eth0/eth1 way) unless REALLY necessary, like in order to make
 older Topspin 3.2 cooperate (this requires eth0/1) with newer Linux like Alma 8.
Doing so requires changes at boot time and compatibility issues may occur later…

OVJ3.2 should work with „new“ network names without issues.
Yet better would be building the latest OVJ via ovjTools as described on Github:

[ovjTools.png]https://github.com/OpenVnmrJ/ovjTools
OpenVnmrJ/ovjTools: Tools and libraries to build OpenVnmrJ. <https://github.com/OpenVnmrJ/ovjTools>
github.com <https://github.com/OpenVnmrJ/ovjTools

Copy/paste instructions are under „Ubuntu and Centos“.

Cheers
Bert


-> Sent via mobile <-

Spin-Doc
Magnetic resonance equipment, service, training & more
Dr. rer. nat. Bert Heise
Owner, applications scientist
Kiefernweg 13
58239 Schwerte
Germany
Web: http://www.spin-doc.net
Mail: bert.heise_at_spin-doc.net
Phone: +49-171-5135315

Representing MestreLab software, Stelar relaxometers & Barthel HF amplifiers

Am 05.08.2026 um 17:20 schrieb Shawn Wagner via groups.io:


Hi Karin,

If you are still having problems, I think I can help you out. I have found
it is best to setup Ubuntu to rename your network card at the startup to eth0
and eth1. This gets rid of the network issues.

If you need, I can send you my SOP for setting up Ubuntu and installing OpenVnmrJ.

Shawn
________________________________
> From: main_at_ammrl.groups.io on behalf of Crowhurst, Karin A via groups.io
> Sent: Friday, July 31, 2026 1:29 PM
> To: main_at_ammrl.groups.io
> Subject: [AMMRL] OpenVnmrJ - problems with linking to NMR console

Hi everyone, I have made it further along in the attempted installation and
setup of OpenVnmrJ to run my Agilent DD2. However, I cannot seem to get the
computer to link with the NMR console. When I open the OpenVnmrJ software,
it says the system

Hi everyone,

I have made it further along in the attempted installation and setup of OpenVnmrJ
to run my Agilent DD2. However, I cannot seem to get the computer to link with
the NMR console. When I open the OpenVnmrJ software, it says the system is
“inactive” (even if I have used the “acqcomm”
command in a terminal to start acquisition communications).

To describe the computer system I am using, it is a Dell XPS, I have installed
Ubuntu 24, and it has one built-in ethernet port, into which I have plugged
the ethernet cable from the NMR, and an adapter ethernet port (plugged in via
USB), to which the ethernet cable for the internet is attached. I have attached
an image of what I see when I use the setacq command. It looks like the ethernet
port for the NMR is not setup. Can anyone provide some insight into how I can
make it communicate?

Many thanks in advance,

Karin Crowhurst

Dr Karin Crowhurst


Professor
Graduate Coordinator

Department of Chemistry and Biochemistry
California State University Northridge
18111 Nordhoff St. MD 8262
Northridge CA 91330-8262
818-677-4288

-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3664): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3664__;!!PvDODwlR4mBZyAb0!UA-CNyltMJYh7C2XqFW6pHkyr0_daZ=
bRCiE_vBHb328_4RVGSc9ZiR-nRe6Ls_Oe0rV2B4fHvKw12ivXyVUOXzjldzh2$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120541079=
/7559972__;!!PvDODwlR4mBZyAb0!UA-CNyltMJYh7C2XqFW6pHkyr0_daZbRCiE_vBHb328_4=
RVGSc9ZiR-nRe6Ls_Oe0rV2B4fHvKw12ivXyVUOX21fyOeS$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Aug 05 2026 - 15:34:53 MST

This archive was generated by hypermail 2.4.0 : Tue Sep 01 2026 - 13:53:27 MST