Re: [AMMRL] OpenVnmrJ - problems with linking to NMR console

From: Shawn Wagner via groups.io <shawn.wagner=usc.edu_at_groups.io>
Date: Wed, 5 Aug 2026 15:15:18 +0000

Hi Karin,

If you are still having problems, I think I can help you out. I have found
it is best to setup Ubuntu to rename your network card at the startup to eth0
and eth1. This gets rid of the network issues.

If you need, I can send you my SOP for setting up Ubuntu and installing OpenVnmrJ.

Shawn
________________________________
> From: main_at_ammrl.groups.io on behalf of Crowhurst, Karin A via groups.io
> Sent: Friday, July 31, 2026 1:29 PM
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: [AMMRL] OpenVnmrJ - problems with linking to NMR console

Hi everyone, I have made it further along in the attempted installation and
setup of OpenVnmrJ to run my Agilent DD2. However, I cannot seem to get the
computer to link with the NMR console. When I open the OpenVnmrJ software,
it says the system


Hi everyone,

I have made it further along in the attempted installation and setup of OpenVnmrJ
to run my Agilent DD2. However, I cannot seem to get the computer to link with
the NMR console. When I open the OpenVnmrJ software, it says the system is “inactive”
(even if I have used the “acqcomm” command in a terminal to start acquisition communications).

To describe the computer system I am using, it is a Dell XPS, I have installed
Ubuntu 24, and it has one built-in ethernet port, into which I have plugged the
ethernet cable from the NMR, and an adapter ethernet port (plugged in via USB),
to which the ethernet cable for the internet is attached. I have attached an image
of what I see when I use the setacq command. It looks like the ethernet port for
the NMR is not setup. Can anyone provide some insight into how I can make it communicate?

Many thanks in advance,

Karin Crowhurst

Dr Karin Crowhurst

Professor
Graduate Coordinator

Department of Chemistry and Biochemistry
California State University Northridge
18111 Nordhoff St. MD 8262
Northridge CA 91330-8262
818-677-4288


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3653): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3653__;!!PvDODwlR4mBZyAb0!WugnMBQ23NJJ152BgalVskKADO_Cmz=
0zI0ARHzLr1RX0U6EQlNUK5uakGH21elzfgSk9yKRdd2QrfqkXK4_2_v_Ukdj6$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120541079=
/7559972__;!!PvDODwlR4mBZyAb0!WugnMBQ23NJJ152BgalVskKADO_Cmz0zI0ARHzLr1RX0U=
6EQlNUK5uakGH21elzfgSk9yKRdd2QrfqkXK4_2_nLVtHkP$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Aug 05 2026 - 08:20:05 MST

This archive was generated by hypermail 2.4.0 : Tue Sep 01 2026 - 13:53:27 MST