Hi,
Just a mini addition to Ken‘s question:
It doesn’t look like a (mechanically) shifted shim coil to me - this should induce
a pronouced asymmetry in the PFG profile (narrower on only one side). Your case looks more like a
change in the parameter-value-to-true-G/cm calibration, leading to a slightly lower gradient
strength than expected/previously, resulting in a generally narrower and taller profile.
Not completely impossible though that the PFG amp output has decreased a bit though - was it
maybe particularly warm in your lab? Some local labs here (yes, in Germany…) unfortunately
aren’t equipped with air conditioning, leading to lall kinds of interesting issues…
Cheers
Bert