Hi all,
I'm troubleshooting a shimming problem on our Bruker Neo400 and would appreciate
any insight from the group.
System and recent history: This magnet regularly cycles between three probes ¡X
an SSNMR iProbe, a Diff BBI probe, and a 5mm solutions iProbe. The recent sequence
was: solids work ¡÷ diffusion work ¡÷ back to the 5mm solutions iProbe.
Shimming was normal through the solids and diffusion sessions; the problems appeared
only after switching back to the 5mm probe.
Symptoms: Since the switch back, I'm seeing persistent shimming issues, including
diverging behavior during automated shimming routines.
Troubleshooting already done:
*
Started shimming from several older, known-good shim sets
*
Redone a full 3D gradient shim from scratch
*
Checked all probe cable/connector seating
*
Ran topshim calreset followed by topshim cal ¡X result: 3 of 8 shims could
not be calibrated, with the software flagging either field instability or a
shim coil problem
Gradient profile anomaly: As part of routine probe health monitoring, I
periodically run gradient profiles to watch for contamination or degradation.
The most recent profile (blue trace) is noticeably narrower and taller than
every historical profile on file for this probe.
Given that everything was behaving normally through the solids and diffusion
sessions, and the issue only surfaced after returning to the 5mm iProbe, I
suspect this is something physical - make the shim stack slipped down a bit?
I'd welcome thoughts on likely causes and next diagnostic steps before I
escalate to a service call.
Thanks in advance,
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3634): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3634__;!!PvDODwlR4mBZyAb0!UxbAmpi7KeGD0gO9rgi_Fam2v65z4k=
XWfHvQWfEUefAr19W6YIoxekYbhXYjWXwFnrzhKjBUCjTSLQenOiYCp565$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120592655=
/7559972__;!!PvDODwlR4mBZyAb0!UxbAmpi7KeGD0gO9rgi_Fam2v65z4kXWfHvQWfEUefAr1=
9W6YIoxekYbhXYjWXwFnrzhKjBUCjTSLQenOq5xb_6a$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Tue Aug 04 2026 - 05:44:42 MST