Hi Hasan

Do you recall during the fill if both helium towers were frosty? And did the
non-filling side frost first? The blockage could be on that side not the
fill side. Another possibility would be a blockage in the transfer line.
That would cause I higher tank pressures but lower exhaust from the magnet.

Blowing helium gas into the magnet is how I have cleared blockages in past.
I have used the standard L-tubes and specialty made things as well. It is
always nerve racking to do this, but can be done. I usually do it right
before the fill.

Greg