 Hi Ken,

I'm not sure if upgrading java is going to do any good.  Topspin has
its own bits of java built-in so I believe it just uses them--doesn't worry
about any other java installation.  

I do see one bug report where it mentions that if you close the edit parameters
popup with the X in the corner you'll see this issue but if you use the OK
button it won't.  It's not exactly similar to what you're seeing though
so I'm not 100% sure that would be a fix.  

Cheers,Mike

    On Saturday, August 16, 2025 at 07:35:36 AM MDT, Kenneth Sharp-Knott via groups.io wrote:

Topspin 4.5.0 on Alma 9

I've been running into a big in ICON where the Edit Parameter buttons stop
working during days of heavy use.  Sort of seems like a memory leak, but
I wonder if upgrading Java might help. I can keep a handle on it by Logging
out of Alma and restarting Topspin/Icon on a daily basis.  Tedious with
6 NMRs and weekends.

Does anyone know the suggested version of Java to use for ICON/Topspin? I
think my Alma 9 instruments are all using OpenJDK v.17 but there is a v.21
out there.  

Has anyone else run into this bug?

Thanks

Ken Sharp-Knott
Manager of Analytical Services and the NMR Facility
Department of Chemistry 
Virginia Tech
(540)267-6502 (Cell)
(540)231-0885 (Office)


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2733): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2733__;!!PvDODwlR4mBZyAb0!R4_F08HJ82w7T6bhVD58YpeTqqXMd6=
ELjXNAimmyDUP58LC9CYxUcb9EA5vRDmVhpMQHxuNgdWIzmYbbNDjWcDMHRzwo$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/114733380=
/7559972__;!!PvDODwlR4mBZyAb0!R4_F08HJ82w7T6bhVD58YpeTqqXMd6ELjXNAimmyDUP58=
LC9CYxUcb9EA5vRDmVhpMQHxuNgdWIzmYbbNDjWcCMc8iVH$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon Aug 18 2025 - 06:31:58 MST

This archive was generated by hypermail 2.4.0 : Mon Sep 01 2025 - 15:37:08 MST