Re: [AMMRL] Stuck sample is BST with SampleMail in the way

From: Kenneth Sharp-Knott via groups.io <kknott=vt.edu_at_groups.io>
Date: Tue, 12 Aug 2025 16:29:46 +0000

Thank you everyone for your advice.

  1.
Yes, I forgot the 5mm BST doesn't have the spinner bit at the bottom. I've
been working with a 5mm BST insert for a wide bore system and on those, the
spinner bit is part of the BST.
  2.
While I did manage to eject the sample eventually, it was clear that all
samples were 'sticking' at the bottom and difficult to eject. So I ended up
warming the prodigy and pulling both it and the shim stack out for
review/cleaning.
  3.
While nothing appeared obviously dirty, It did notice a little fleck of
something come out with the probe... Might have been something had gotten
into the bore that was causing the issue.
  4.
After cleaning and reassembly, samples are spinning and ejecting as normal.

Now we wait until tomorrow for the cooldown to complete and begin the 3D
shim routine.

Thanks again.

Ken Sharp-Knott
Manager of Analytical Services and the NMR Facility
Department of Chemistry
Virginia Tech

(540)267-6502 (Cell)
(540)231-0885 (Office)
________________________________
> From: main_at_ammrl.groups.io on behalf of Alex Greenwood via groups.io
> Sent: Tuesday, August 12, 2025 11:55 AM
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: Re: [AMMRL] Stuck sample is BST with SampleMail in the way

Hi Ken,

If this turns out to not be an isolated incident, you may want to clean the
top of your shim stack, both the beveled part on top where the spinner rests
and an inch or so below that the lower part of the spinner inserts into.
Dirt in these areas can cause spinning issues, prevent the sample from
inserting all the way (leading to lock/shim failures) and prevent the sample
from ejecting reliably. I started having these issues on two systems earlier
this year and in both cases cleaning the shim stack (water/isopropanol) fixed
the issue. In my case it was a matter of removing the probe, removing the
autosampler to access/loosen the three long screws that hold the BST and
shim stack together, then loosening the clamp under the magnet that holds
onto the shims.

Alex


Alex Greenwood (he/him)
Manager, NMR Facility
University of Cincinnati
123C Crosley Hall
Office: 513-556-9211; Cell: 412-952-8044
https://www.artsci.uc.edu/natural-sciences/chemistry/core-facilities/nuclear-magnetic-resonance-facility.html



> From: main_at_ammrl.groups.io On Behalf Of Steve Clemens via groups.io
> Sent: Tuesday, August 12, 2025 11:03 AM
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] Stuck sample is BST with SampleMail in the way

Ken,

Depending on how the sample is "stuck", you might not need to do any of this.
I have had a situation where a sample was in the bore but was binding against
something and wouldn't come up. With the lift gas on, I lowered a stick
down (with a small ball of kimwipe on the end) just to 'touch' the top of
the sample, at which point it righted itself and came up.

I suggest this as a really easy way to deal with the issue.

Also, if I understand the situation correctly, I personally would be very
wary of removing the BST with the probe still mounted. The bottom part of
the BST (where it mates to the shim stack/turbine) is the same diameter as
the outside of the upper part of the spinner, so this could only help if the
sticking is between the BST and the outer part of the spinner (ie. where the
reflective tape is). Otherwise you will pull the BST up and the sample
will still be sitting on top of the shim stack/turbine part, in which case
you'll need to drop your probe (and remove the shim stack) to get that out
anyway.

Also, if it is stuck on the side of the BST, there is a risk of it coming
loose as you pull the BST up, and then you'll be cleaning up a mess (as it
will fall back down the bore and almost certainly break the tube).

Steve

-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2715): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2715__;!!PvDODwlR4mBZyAb0!RgPKJBugGk2xLuAhd_0CiDmsUzGmrj=
H__Rr6b65M0qYT7SFo742_f5X4K9CZBOZjEzOC22zDNDaPqGmbFaRVsACs$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/114652860=
/7559972__;!!PvDODwlR4mBZyAb0!RgPKJBugGk2xLuAhd_0CiDmsUzGmrjH__Rr6b65M0qYT7=
SFo742_f5X4K9CZBOZjEzOC22zDNDaPqGmbFYS7exsK$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Aug 12 2025 - 09:31:03 MST

This archive was generated by hypermail 2.4.0 : Mon Sep 01 2025 - 15:37:08 MST