Hi Ken

unfortunately there is no simple way as the sampleMail/Case is connected with
the BST. There are a couple of screws connecting the horizontal arm with the
BST. Once these are out you can usually push it out of the way or you loosen
a few more screws at the horizontal brace. Then you get to the 3 screws
that connect the BST to shims. When they are unscrewed the shims will drop
down as everything is anchored at the top.

You might get some advice form the community to use a broomstick with putty
at the end to fish a sample out. If that works it is a lot easier.

Regards

Clemens


On 8/11/2025 2:33 PM, Kenneth Sharp-Knott via groups.io wrote:

I have a sample that seems to be stuck in the BST.� I can't remove
the BST without moving the SampleMail portion of the SampleCase.�
Unfortunately, I think I skipped class when they went over this....

How do you remove the BST when you have a Sample Case/Mail system in place?

Thanks,

Ken Sharp-Knott
Manager of Analytical Services and the NMR Facility
Department of Chemistry�
Virginia Tech

(540)267-6502 (Cell)
(540)231-0885 (Office)



--
==========================
==========================
====================
Clemens Anklin Ph.D. Vice President
Bruker BioSpin NMR Applications & Training
15 Fortune Drive NEW Phone: 978-313-5NMR(5667)
Billerica MA 01821 bluesky: _at_canklin.bsky.social
web: http://www.bruker.com
e-mail: clemens.anklin_at_bruker.com<
==========================
==========================
====================



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2705): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2705__;!!PvDODwlR4mBZyAb0!TLa22iPiwnIrPuQRKVDNaS3zPuRKV9=
pRh_RxxVtbhbCgX9hMVlfqh8m5II9e6oi0ZFwcLNDG-L0yrP5Mg9RDtmj4pAUZ$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/114652860=
/7559972__;!!PvDODwlR4mBZyAb0!TLa22iPiwnIrPuQRKVDNaS3zPuRKV9pRh_RxxVtbhbCgX=
9hMVlfqh8m5II9e6oi0ZFwcLNDG-L0yrP5Mg9RDtm0RcikT$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon Aug 11 2025 - 12:03:19 MST

This archive was generated by hypermail 2.4.0 : Mon Sep 01 2025 - 15:37:08 MST