Re: [AMMRL] Agilent/Varian SMS Lift, Reach, and Rotation Potentiometer Replacement

From: Spin-Doc <bert.heise_at_spin-doc.net>
Date: Sun, 10 Aug 2025 16:40:23 +0200

Hi Laird,

I have so far bought these Bourns potentiometers from Mouser, part number
652-3549H-1AF-103A

They have a tendency of being a tiny bit (too) thick so it's a bit of a squeeze
to get it in, but it does work.

They come with that aluminium plate/disk installed, so they were not added
later.

Cheers
Bert


> Am 09.08.2025 um 23:48 schrieb Laird Trimble:
>
> Hi All,
>
> I've got a couple of old Agilent/Varian Zymark sample changers and it is
> getting time to replace some of the potentiometers used for sensing lift,
> reach, and rotation positions. I have the following official unused Agilent
> replacement parts:
>
> Lift: Bourns Hybritron 3541H-471-103L RES 10Kohm +/- 10%, LIN +/-0.1% (Agilent Part No. 1901089800)
> Reach: Bourns Hybritron 3541H-470-103L RES 10Kohm +/- 10%, LIN +/-0.1% (Agilent Part No. 1901099200)
> Rotation: Bourns Hybritron 3541H-472-103L RES 10Kohm +/- 10%, LIN +/-0.1% (Agilent Part No. 1901089900)
>
> All three pots appear to be 10Kohm ±10% with a ±0.1% linearity.
> And yet they have slightly different Bourns and Agilent part numbers. As
> far as I can tell electrically they are the same part. I am curious, are
> they the same? Will any 10 turn linear 10Kohm pot with the same linearity
> be a suitable replacement?
>
> Several SMS robots that I have opened have pots manufactured by ETI Systems,
> part number MH22S-10. These pots have the same electrical characteristics
> and linearity as the Bourns part but the ETI parts used at all three
> locations have the exact same part number. This leads me to think that
> all three Bourns pots are essentially the same thing.
>0
> Finally, the pots used in the SMS have a disk that is attached to the body
> that is used to secure the pot to the SMS housing rather than the screw
> type mount that is more common for this type of pot. Does anyone know if
> this ring is installed by Agilent/Varian or by the original manufacturers?
>
> Thanks all for your help.
>


Magnetic resonance equipment, service, training & more
Selling & buying used NMR equipment (all vendors)
Dr. rer. nat. Bert Heise
Owner, Applications Scientist

Spin-Doc Bert Heise
Kiefernweg 13
58239 Schwerte
Germany
Mobile: +49 171 5135315
Email: bert.heise_at_spin-doc.net
Web: http://www.spin-doc.net

Representing:
Stelar Fast Field Cycling NMR Relaxometers (http://www.stelar.it)
MestreLab NMR Software (http://www.mestrelab.com)
Barthel HF Amplifiers (http://www.barthel-hf.de_)



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2701): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2701__;!!PvDODwlR4mBZyAb0!TyuVx3yV6HCwWFB2u9TX6dPRcS6mde=
VD7LjVwxpUoqLsJZUlst1qjna3tSuYidaiHrMdgerB0_d5T2CBSz0SMA$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/114629793=
/7559972__;!!PvDODwlR4mBZyAb0!TyuVx3yV6HCwWFB2u9TX6dPRcS6mdeVD7LjVwxpUoqLsJ=
ZUlst1qjna3tSuYidaiHrMdgerB0_d5T2D-jMIfEA$
Group Owner: main+owner_at_ammrl.groups.io
Unsubscribe: https://urldefense.com/v3/__https://ammrl.groups.io/g/main/lea=
ve/12177410/7559972/1125599165/xyzzy__;!!PvDODwlR4mBZyAb0!TyuVx3yV6HCwWFB2u=
9TX6dPRcS6mdeVD7LjVwxpUoqLsJZUlst1qjna3tSuYidaiHrMdgerB0_d5T2CfxSDyJw$ [wa=
lt_at_hawaii.edu]
-=-=-=-=-=-=-=-=-=-=-=-



Received on Sun Aug 10 2025 - 09:00:12 MST

This archive was generated by hypermail 2.4.0 : Mon Sep 01 2025 - 15:37:08 MST