Re: [AMMRL] Nitrogen sensor fluctuating on AS600 magnet

From: Holmes, Daniel via groups.io <holmesd5=msu.edu_at_groups.io>
Date: Tue, 5 Aug 2025 17:26:53 +0000

I had the same issue on the same magnet. It turned out to be a faulty/loose
connection at the sensor input on the magnet. Nothing more than a little
jiggle and it was normal again.

Dan



-=-=-=-=-=-=-=-=-=-=-=-