Re: [AMMRL] AVIII lock lost problem.

From: Julian Holzinger via groups.io <Julian.Holzinger=cup.lmu.de_at_groups.io>
Date: Fri, 1 Aug 2025 10:09:27 +0200

Dear Tianzhi,

as Jay already mentioned, it might be a problem with the temperature.
We had it once at a AvanceIIIHD SingleBay that several fans of the BSMS
were malfunctioning and the lock shut down by itself (the light at L-TRX
turned red) - upon waiting for a while without running an experiment it
came back to life. The problem was that fan failure would've been only
reported for the fans in the AQS unit but not in the BSMS. You can check
the fans with a piece of paper or a plastic cable tie.

Good luck and all the best
Julian



Am 31/07/2025 um 19:17 schrieb Wang, Tianzhi via groups.io:
> Dear All
>
> I was experiencing an intermittent lock loss issue on my AVIII 800 MHz
> NMR spectrometer with a cryoprobe (CP). The lock was lost and then
> came back automatically when experiments were running or idle
> recently. It looks like it became better after I rebooted the BSMS
> unit, but it happened again after 2-3 weeks. My best guess is that the
> L-TRX board was not stable, but I don't have a spare L-TRX board to
> test. I checked all the cables and connectors which are fine.  Is
> there anyone who can share the experience to solve such problem with me?
>
> Best,
> Tianzhi
>
> -------------------------------------------------------------------
> Tianzhi Wang, Ph. D.
> NMR Manager
> Sealy center for Structural Biology & Molecular Biophysics
> University of Texas Medical Branch,
> 301 University Bvld, NMR dockside building, Rom 1.104
> Galveston, TX 77555-1157
> Email: tiawang_at_utmb.edu <mailto:tiawang_at_utmb.edu>
> (409)7476821 (O)
>
>


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2652): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2652__;!!PvDODwlR4mBZyAb0!QVAzWPuOdUYI1_nidokErvYFak391f=
2aD68ehLCEKcPXxL1DCFNIbl8Tk2C6H7Jk0HqrVVBg8O4PJrZDUCQadq5pCbos$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/114470642=
/7559972__;!!PvDODwlR4mBZyAb0!QVAzWPuOdUYI1_nidokErvYFak391f2aD68ehLCEKcPXx=
L1DCFNIbl8Tk2C6H7Jk0HqrVVBg8O4PJrZDUCQadoXnaPhw$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri Aug 01 2025 - 08:26:46 MST

This archive was generated by hypermail 2.4.0 : Mon Sep 01 2025 - 15:37:08 MST