Re: [AMMRL] Off axis shims bad after 24 hours on 40-year-old Oxford magnet #Bruker #Hardware

From: Samrat Amin <saamin1_at_mainex1.asu.edu>
Date: Sat, 17 Aug 2024 13:06:27 -0700

Hi All,

I can't offer any advice to the original question, but I'll throw in some
information that can hopefully benefit others here.  As many have pointed
out in this thread, its really helpful to monitor the magnets and have long
term data to compare.   Alongside our helium recovery system, I
installed some sensors (with custom electronics) to monitor everything about
our magnets and environment, and the data has been incredibly useful for
troubleshooting various events in our operations.  It was a LOT of
effort to setup/design/program, but worth it for the ability to correlate
events and monitor the performance of instruments over time.

My system monitors helium boiloff flow (we can add LN2 flow as well), line
pressures, building gas pressures, heat exchanger temperatures, vacuum pump
temperatures, and basically all environmental parameters (temp, humidity,
pressure, etc).  All the data is saved to a database and viewed online
in real time on a dashboard.  It has also been setup to send alerts via
slack if any problems arise (its definitely saved us a few times already
when equipment misbehaves!).

For those that are interested in seeing it in action, here is a public link
to our dashboard (I'll keep it public for a week or so, or until IT gets upset):
https://bit.ly/4dt7ZPD

I would highly recommend a similar kind of setup that allows you to log data
long term.  If nothing else, monitoring the helium/LN2 boiloff will at
least give some sense of the health of the magnets.   Our sensors
are sensitive enough to see the small changes in boiloff when you measure
the helium level (you can notice the small periodic spikes in the data).

I apologize for the tangent from the original question.  But if this kind
of thing is of interest among the NMR community, I am happy to setup a separate
thread with more details on setup.

Wendy - if you think you're headed towards a quench, it may be helpful to
monitor the helium flow and whether its increasing over time.   I'd
be happy to send you one of our old sensors that monitors helium boiloff if
you think its useful.

Best,

Samrat Amin
Magnetic Resonance Research Center
Arizona State University


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1530): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1530__;!!PvDODwlR4mBZyAb0!XL-Q_TTNnoWsFScmixPHzue1e--Tgm=
a2dTAyFs7SI3YzTzxSPD8Ywtmj0xgl6xy8Qltw7JU11NQILO5JAI7wiA$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/107818605=
/7559972__;!!PvDODwlR4mBZyAb0!XL-Q_TTNnoWsFScmixPHzue1e--Tgma2dTAyFs7SI3YzT=
zxSPD8Ywtmj0xgl6xy8Qltw7JU11NQILO4vKM_z-Q$
Mute #bruker:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/bruker__;!!PvDODwlR4mBZyAb0!XL-Q_TTNnoWsFScmixPHzue1e--Tgma2dTAyFs=
7SI3YzTzxSPD8Ywtmj0xgl6xy8Qltw7JU11NQILO6Nt1k-jQ$
Mute #hardware:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/m=
utehashtag/hardware__;!!PvDODwlR4mBZyAb0!XL-Q_TTNnoWsFScmixPHzue1e--Tgma2dT=
AyFs7SI3YzTzxSPD8Ywtmj0xgl6xy8Qltw7JU11NQILO5Y0j8Ltw$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Sat Aug 17 2024 - 13:06:30 MST

This archive was generated by hypermail 2.4.0 : Sun Sep 01 2024 - 15:22:33 MST