Hello Tom,

We have exactly the same problem with our AV3-400. So far we found out the
following:

Depending on the selection of your solvent topshim will use 1H or 2H. So for
Topshim 3D, where you use a 10%D2O/H2O solvent tosphim will pulse on 1H.
For Topshim 2D, using D2O, Topshim will pulse on 2H. This explains your
observations: the system is not sending 2H rf pulses but 1H is ok.

In our case, we have tried repeatedly a reset of all components without much
success. Pulses on the 2H channel are generated in the 2H/TX board.
We measured the output of this board using an oscilloscope during pulsing on
2H and output much too low. Hence the console is not sending 2H pulses to
the lock-coil during topshim.

Alternatively, you can also check the 2H 90 degree pulse length when pulsing
on the lock channel.   We have ordered a replacement for our 2H/Tx
board and I will let you know how this works out.    You should also
check the tuning of the lock-channel of your probe.

Kind Regards,

Hanno
Georgia Tech


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1523): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1523__;!!PvDODwlR4mBZyAb0!Vxtm5TQnBshbik85VkQVFPTHicXCA4=
-3yc-ArCQx4tmt6PG7CJ2f8C9X6jK7scY6lkcChP7Znttu96k$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/107914803=
/7559972__;!!PvDODwlR4mBZyAb0!Vxtm5TQnBshbik85VkQVFPTHicXCA4-3yc-ArCQx4tmt6=
PG7CJ2f8C9X6jK7scY6lkcChP7Z1MQILaI$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Aug 15 2024 - 16:09:18 MST

This archive was generated by hypermail 2.4.0 : Sun Sep 01 2024 - 15:22:33 MST