Re: [AMMRL] Off axis shims bad after 24 hours on 40-year-old Oxford magnet #Bruker #Hardware

From: Spin-Doc <bert.heise_at_spin-doc.net>
Date: Thu, 15 Aug 2024 22:53:51 +0200
Hi,

Could you possibly specify “lineshape deterioration” a bit more precisely? I mean, are we talking about a change from 1 Hz linewidth at half-height to something like 5-10 Hz (not great, but can easily be induced by e.g. changes in the sample like bubble formation, product fallout or solvent evaporation) or 200-2000 Hz (making up numbers, but that’s what a failing cryoshim can easily lead to)?

A failing cryoshim is also one-way only, so lineshape is and stays bad afterwards and an only rarely be compensated by RT shimming. Your case doesn’t sound like a cryoshim effect to me…

Best regards 
Bert